Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Tropo...

FDA Device Recall #Z-1025-2026 - Class II, November 26, 2025

Recall Summary

Recall Number Z-1025-2026
Classification Class II - Moderate risk
Date Initiated November 26, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Beckman Coulter, Inc.
Location Chaska, MN
Product Type Devices
Quantity 1272 units

Product Description

Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit

Reason for Recall

Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.

Distribution Pattern

Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark, Dominican Republic, Egypt, El Salvador, Eswatini, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Moldova, Republic of, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam, and Zimbabwe.

Lot / Code Information

A71460: UDI 15099590369224 A71461: UDI 15099590369248 A30260: UDI 15099590341602

Other Recalls from Beckman Coulter, Inc.

Recall # Classification Product Date
Z-1922-2026 Class II Access Total T4 Calibrator, Catalog No. 33805 Mar 16, 2026
Z-1235-2026 Class II CellMek SPS Sample Preparation System, REF: C44... Dec 23, 2025
Z-1026-2026 Class II Beckman Coulter UniCel Dxl 800, Part numbers: ... Nov 26, 2025
Z-1221-2026 Class II COULTER DxH Diluent, REF 628017 COULTER DxH EC... Nov 10, 2025
Z-0889-2026 Class II Access 2 Reaction Vessels, individual, disposab... Nov 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.