CellMek SPS Sample Preparation System, REF: C44603, with software

FDA Recall #Z-1235-2026 — Class II — December 23, 2025

Recall #Z-1235-2026 Date: December 23, 2025 Classification: Class II Status: Ongoing

Product Description

CellMek SPS Sample Preparation System, REF: C44603, with software

Reason for Recall

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Recalling Firm

Beckman Coulter, Inc. — Miami, FL

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

50

Distribution

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

Code Information

UDI-DI: 15099590750312. Software Version 2.3.90.0

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated