EndoWrist Stapler 45 and Stapler 30 instruments used on the da Vinci Xi systems (IS4000) with p...

FDA Device Recall #Z-1568-2016 — Class II — March 25, 2016

Recall Summary

Recall Number Z-1568-2016
Classification Class II — Moderate risk
Date Initiated March 25, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Intuitive Surgical, Inc.
Location Sunnyvale, CA
Product Type Devices
Quantity 43

Product Description

EndoWrist Stapler 45 and Stapler 30 instruments used on the da Vinci Xi systems (IS4000) with p5 software. General Surgery: The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.

Reason for Recall

Potential for unexpected motion of the Xi Stapler jaws on the da Vinci Xi System with p5 software relating to a combination of the p5 software and the surgeon quickly transitioning from the clamp to the fire pedal during use.

Distribution Pattern

US Nationwide Distribution in the states of: WI, FL, OH, TX, CA, AZ, IL, MD, MN, NY, OK, HI, MO, IA, KS, PA, and AL.

Lot / Code Information

da Vinci systems with serial numbers: SK0036 SK0292 SK0650 SK0039 SK0293 SK0652 SK0057 SK0294 SK0653 SK0101 SK0326 SK0654 SK0106 SK0358 SK0665 SK0119 SK0360 SK0701 SK0143 SK0382 SK0702 SK0144 SK0408 SK0710 SK0146 SK0438 SK0722 SK0171 SK0456 SK0728 SK0173 SK0484 SK0181 SK0492 SK0216 SK0496 SK0230 SK0516 SK0233 SK0548 SK0237 SK0552 SK0238 SK0610 SK0263 SK0612 SK0286 SK0631 SK0291 SK0649.

Other Recalls from Intuitive Surgical, Inc.

Recall # Classification Product Date
Z-1355-2026 Class II da Vinci 5 Surgeon Console Viewer Display, ASSY... Dec 30, 2025
Z-1096-2026 Class II da Vinci SP, SP Access Port Kit, Large Incisio... Dec 15, 2025
Z-1134-2026 Class II Brand Name: Ion Endoluminal System Product Nam... Dec 4, 2025
Z-1428-2026 Class II Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... Nov 17, 2025
Z-1232-2026 Class II Brand Name: Da Vinci Product Name: da Vinci X,... Nov 7, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.