Date Product Reason Class
Sep 11, 2025 ECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection... The devices were shipped unsterilized. Class II
Apr 22, 2025 Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 ro... Firm has received an increase in reports regarding inadvertent instrument loc... Class I
Dec 11, 2023 ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON... There is the potential for incomplete staple line formation during the firing... Class II
Jul 20, 2021 HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft tissue incisi... An internal component may be cracked and become lodged behind the energy butt... Class II
Jul 20, 2021 HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisio... An internal component may be cracked and become lodged behind the energy butt... Class II
Jun 3, 2021 ECHELON FLEX Powered Plus Stapler-intended for transection, resection, and/or... Mislabeled: Labeled on the package as 60mm contains 45mm devices Class II
Nov 7, 2019 Ligamax 5 mm Endoscopic Multiple Clip Applier A potential exists that certain LIGAMAX devices within the impacted lots may... Class II
Oct 3, 2019 ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code P... The staplers may contain an out of specification anvil component within the j... Class I
Oct 3, 2019 ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 340mm shaft, Product Code P... The staplers may contain an out of specification anvil component within the j... Class I
Oct 3, 2019 ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code P... The staplers may contain an out of specification anvil component within the j... Class I
Oct 3, 2019 ECHELON FLEX ENDOPATH 60mm Stapler - 340mm shaft, Product Code EC60A The staplers may contain an out of specification anvil component within the j... Class I
May 15, 2019 Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 25cm Shaft PRODUCT CODE: NSLX125C Observed reported incidence rate for harm associated with intraoperative blee... Class II
May 15, 2019 Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 37cm Shaft PRODUCT CODE: NSLX137C Observed reported incidence rate for harm associated with intraoperative ble... Class II
May 15, 2019 Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 45cm Shaft PRODUCT CODE: NSLX145C Observed reported incidence rate for harm associated with intraoperative ble... Class II
Apr 11, 2019 Curved Intraluminal Staplers, 33 mm diameter, Model CDH33A The Ethicon End... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Endoscopic Curved Intraluminal Stapler, 29 mm diameter, Model ECS29A The E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Curved Intraluminal Staplers, 25 mm diameter, Model CDH25A The Ethicon End... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Curved Intraluminal Staplers, 29 mm diameter, Model CDH29A The Ethicon End... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Straight Intraluminal Staplers, 21 mm diameter, Model SDH21A The Ethicon E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Endoscopic Curved Intraluminal Stapler, 25 mm diameter, Model ECS25A The E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Curved Intraluminal Staplers, 21 mm diameter, Model CDH21A The Ethicon End... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Straight Intraluminal Staplers, 29 mm diameter, Model SDH29A The Ethicon E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Endoscopic Curved Intraluminal Stapler, 33 mm diameter, Model ECS33A The E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Endoscopic Curved Intraluminal Stapler, 21 mm diameter, Model ECS21A The E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Straight Intraluminal Staplers, 33 mm diameter, Model SDH33A The Ethicon E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Apr 11, 2019 Straight Intraluminal Staplers, 25 mm diameter, Model SDH25A The Ethicon E... The staplers may have an insufficient firing stroke to break the washer and c... Class I
Dec 7, 2018 PROXIMATE¿ HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: ... Hemorrhoidal Circular Stapler may have been assembled without a washer, this ... Class II
Dec 7, 2018 PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: Th... Hemorrhoidal Circular Stapler may have been assembled without a washer, this ... Class II
May 3, 2018 HARMONIC ACE Shears + Adaptive Tissue Technology, 5mm Diameter / 23cm Length ... The devices may have been assembled with an internal component that may cause... Class II
May 3, 2018 HARMONIC ACE Shears + Adaptive Tissue Technology, 5mm Diameter / 36cm Length ... The devices may have been assembled with an internal component that may cause... Class II
Mar 26, 2018 5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability S... Use of the product with insufficient lubrication may prolong surgery due to t... Class II
Mar 26, 2018 5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability S... Use of the product with insufficient lubrication may prolong surgery due to t... Class II
Mar 26, 2018 5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability S... Use of the product with insufficient lubrication may prolong surgery due to t... Class II
Mar 26, 2018 5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability S... Use of the product with insufficient lubrication may prolong surgery due to t... Class II
Mar 26, 2018 5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability S... Use of the product with insufficient lubrication may prolong surgery due to t... Class II
Oct 23, 2017 Disposable Curved Circular Staplers Product Usage: Staplers have applicat... There is a potential for the devices to create staples that do not meet forme... Class II
Mar 31, 2017 Endopath ETS-FLEX Endoscopic Articulating Linear Cutter, VASCULAR/THIN 35MM There is a risk that the pinion gear in the device could fail under extreme u... Class II
Jul 15, 2016 Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOP... Quality control identified a component issue where the pinion gear in the dev... Class II
Jul 15, 2016 Endoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH E... Quality control identified a component issue where the pinion gear in the dev... Class II
Sep 23, 2015 HARMONIC ACE+ 7 Laparoscopic Shears with Advanced Hemostasis (23cm, 36cm, 45c... Complaints of holes in Tyvek which can compromise the sterile barrier provide... Class II
May 13, 2015 SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An inte... Ethicon has found that disinfecting methods not specified in the Instructions... Class I
Feb 26, 2015 ENDOPATH ECHELON" Vascular White Reload for Advanced Placement Tip, Produ... During an internal inspection of the production process the firm discovered a... Class II
Aug 12, 2014 PROXIMATE Skin Stapler. PROXIMATE RH Rotating-Head Skin Stapler (Model PRR3... The device may have a non-conforming component that may cause the device to f... Class II
Sep 30, 2013 Adjustable Gastric Band, packaged alone in transparent PETG inner and outer b... During aging studies, the firm determined that the gastric belts were unable ... Class II
Sep 30, 2013 Adjustable Gastric Band, packaged alone in transparent PETG inner and outer b... During aging studies, the firm determined that the gastric belts were unable ... Class II
Sep 30, 2013 Adjustable Gastric Band, packaged alone in transparent PETG inner and outer b... During aging studies, the firm determined that the gastric belts were unable ... Class II
May 13, 2013 ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack ... Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 6... Class II
Apr 12, 2013 LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L, bulk n... Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L ... Class II
Apr 12, 2013 LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier Convenience Kit (Flex Tra... Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L ... Class II
Apr 12, 2013 LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Pro... Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L ... Class II

View all 54 device recalls →

Frequently Asked Questions

A high number of recalls does not necessarily indicate that a company is unsafe. Large manufacturers that produce thousands of products across multiple categories will statistically appear in recall databases more frequently. What matters more is the severity of each recall (Class I being the most serious), the speed of response, and whether the company proactively identified and addressed the issue. Companies with robust safety programs often catch problems earlier.

Food, drug, and medical device recall data comes from the FDA's openFDA enforcement database, which contains all FDA enforcement reports. Consumer product recall data comes from the U.S. Consumer Product Safety Commission (CPSC). Both sources are public government databases that are updated regularly. RecallCheck aggregates these sources to provide a unified view of a company's recall history.

The FDA classifies recalls into three categories. Class I is the most serious — there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. Class II means the product may cause temporary or medically reversible health problems, or the probability of serious consequences is remote. Class III is the least serious — the product is unlikely to cause adverse health consequences. CPSC product recalls do not use this classification system.

Yes. You can search for any company using the firm/manufacturer pages for each category: Food Firms, Drug Firms, Device Firms, or Product Manufacturers. You can also use the search functionality on any browse page to find recalls by company name.

Explore More Recalls

Search recalls by category, state, reason, or firm across all our databases.