PROXIMATE¿ HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE¿ HCS H...
FDA Recall #Z-0695-2019 — Class II — December 7, 2018
Product Description
PROXIMATE¿ HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE¿ HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.
Reason for Recall
Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired
Recalling Firm
Ethicon Endo-Surgery Inc — Blue Ash, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
480 units
Distribution
Worldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.
Code Information
Lot Codes: P9461W, P94K20
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.