Xoran MiniCAT00202
FDA Device Recall #Z-0040-2014 - Class II, July 27, 2010
Recall Summary
| Recall Number | Z-0040-2014 |
| Classification | Class II - Moderate risk |
| Date Initiated | July 27, 2010 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Xoran Technologies, Inc. |
| Location | Ann Arbor, MI |
| Product Type | Devices |
| Quantity | 320 |
Product Description
Xoran MiniCAT00202
Reason for Recall
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.
Distribution Pattern
Worldwide Distribution - Nationwide
Lot / Code Information
minicat
Other Recalls from Xoran Technologies, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0039-2014 | Class II | Xoran Minicat00158 | Jul 27, 2010 |
| Z-0037-2014 | Class II | Xoran MiniCAT CT Scanner | Jul 27, 2010 |
| Z-0038-2014 | Class II | Xoran Minicat00202 | Jul 27, 2010 |
| Z-0036-2014 | Class II | Xoran Mini CT Scanner | Jul 27, 2010 |
| Z-0041-2014 | Class II | Xoran MiniCAT CT Scanner ID ENT00001 | Dec 14, 2009 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.