VITROS XT 3400 Chemistry System, Catalog No. 6844458

FDA Device Recall #Z-1232-2024 — Class II — January 26, 2024

Recall Summary

Recall Number Z-1232-2024
Classification Class II — Moderate risk
Date Initiated January 26, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Ortho-Clinical Diagnostics, Inc.
Location Rochester, NY
Product Type Devices
Quantity 25 US; 26 OUS

Product Description

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Reason for Recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Lot / Code Information

VITROS System Software Versions 3.8.0 or 3.8.1 UDI-DI 10758750031986 Serial Numbers: US: J34500140 J34500180 J34500257 J34500289 J34500319 J34500386 J34500422 J34500447 J34500473 J34500474 J34500510 J34500539 J34500547 J34500579 J34500588 J34500622 J34500656 J34500777 J34500818 J34500861 J34500887 J34500893 J34500896 J34500911 J34500931 OUS: J34500123 J34500128 J34500129 J34500130 J34500190 J34500261 J34500269 J34500271 J34500292 J34500301 J34500337 J34500356 J34500363 J34500515 J34500520 J34500536 J34500582 J34500609 J34500628 J34500658 J34500711 J34500793 J34500802 J34500940 J34500960 J34500962

Other Recalls from Ortho-Clinical Diagnostics, Inc.

Recall # Classification Product Date
Z-2559-2025 Class II Chemistry Products LAC Slides. Catalog Number: ... Jul 29, 2025
Z-2560-2025 Class II VITROS Chemistry Products LAC Slides. Catalog N... Jul 29, 2025
Z-2335-2025 Class II VITROS Chemistry Products Ca Slides; Catalog N... Jul 9, 2025
Z-2073-2025 Class II VITROS Chemistry Products OP Reagent Gen 15. Mo... Jun 12, 2025
Z-2121-2025 Class II Brand Name: VITROS Chemistry Products Product ... Jun 10, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.