VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
FDA Device Recall #Z-3231-2026 - Class II, July 24, 2026
Recall Summary
| Recall Number | Z-3231-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | July 24, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Biomerieux Inc |
| Location | Hazelwood, MO |
| Product Type | Devices |
| Quantity | 97680 units |
Product Description
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Reason for Recall
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Distribution Pattern
International distribution to the countries of Austria, Germany, Luxembourg.
Lot / Code Information
UDI-DI 03573026637712; Lot Number 1333393504
Other Recalls from Biomerieux Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2715-2026 | Class II | VITEK 2 AST-N447 TEST KIT Product reference 42... | Jun 4, 2026 |
| Z-1236-2026 | Class II | VITEK 2 Gram-negative Susceptibility Cards Cont... | Jan 5, 2026 |
| Z-1464-2024 | Class I | VITEK 2 - AST Test kits (in vitro diagnostic te... | Mar 14, 2024 |
| Z-1201-2023 | Class II | MYLA comprises AST Filters in conjunction with ... | Feb 6, 2023 |
| Z-0798-2023 | Class II | VIDAS CLINICAL VIDAS LISTERIA 60T, CATALOG 30700 | Nov 9, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.