VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

FDA Device Recall #Z-0073-2027 - Class II, August 31, 2026

Recall Summary

Recall Number Z-0073-2027
Classification Class II - Moderate risk
Date Initiated August 31, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Steris Corporation
Location Mentor, OH
Product Type Devices
Quantity 2205

Product Description

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

Reason for Recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Distribution Pattern

Worldwide - US Nationwide distribution.

Lot / Code Information

UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027

Other Recalls from Steris Corporation

Recall # Classification Product Date
Z-0118-2027 Class II HexaVue Focus Integration System, a single, all... Sep 23, 2026
Z-0075-2027 Class II VERIFY Biological Steam Test Pack Model/Catalo... Aug 31, 2026
Z-0070-2027 Class II VERIFY Biological/Integrator Challenge Pack ... Aug 31, 2026
Z-0074-2027 Class II VERIFY Biological Steam Test Pack Model/Catalo... Aug 31, 2026
Z-0071-2027 Class II VERIFY Biological/Integrator Challenge Pack ... Aug 31, 2026

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately, you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.