The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit b...
FDA Device Recall #Z-0963-2026 - Class II, November 12, 2025
Recall Summary
| Recall Number | Z-0963-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | November 12, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Howmedica Osteonics Corp. |
| Location | Mahwah, NJ |
| Product Type | Devices |
| Quantity | 320 units |
Product Description
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Reason for Recall
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Lot / Code Information
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Other Recalls from Howmedica Osteonics Corp.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2461-2026 | Class II | TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300. | May 19, 2026 |
| Z-2462-2026 | Class II | TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400. | May 19, 2026 |
| Z-1572-2026 | Class II | MICS3 Angled Sagittal Saw Attachment; Part Num... | Feb 2, 2026 |
| Z-2640-2025 | Class II | 1. EXETER V40 STEM 44MM NO 2, Model/Catalog Num... | Aug 28, 2025 |
| Z-0023-2025 | Class II | Triathlon Universal TS Baseplate Size 2 - A sin... | Sep 4, 2024 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.