The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated...
FDA Device Recall #Z-1415-2021 - Class II, December 22, 2020
Recall Summary
| Recall Number | Z-1415-2021 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 22, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | EOS Imaging |
| Location | Paris, N/A |
| Product Type | Devices |
| Quantity | 11 devices |
Product Description
The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. ¿ The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). ¿ The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.
Reason for Recall
Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
Distribution Pattern
Distributed nationwide to MN, OH and internationally to Australia, Canada, Germany, Spain, and France
Lot / Code Information
Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.1909.1006, 8.2005.1014, 8.1905.1004, 8.1911.1007, 8.1912.1009, and 8.2006.1016
Other Recalls from EOS Imaging
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-3188-2018 | Class II | sterEOS workstation | Aug 1, 2018 |
| Z-2626-2017 | Class II | ROTH MOBELI Grab Handles used with EOS System: ... | May 30, 2017 |
| Z-0580-2014 | Class II | steroEOS Workstation with software versions: 1.... | Nov 29, 2013 |
| Z-0529-2014 | Class II | EOS System X- ray beam Digital radiography s... | Nov 7, 2013 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.