Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, RE...
FDA Device Recall #Z-2888-2016 - Class II, December 29, 2009
Recall Summary
| Recall Number | Z-2888-2016 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 29, 2009 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Spiegelberg Gmbh & Co. KG |
| Location | Hamburg, N/A |
| Product Type | Devices |
| Quantity | 22 units |
Product Description
Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Reason for Recall
In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
Distribution Pattern
Nationwide Distribution to PA only
Lot / Code Information
Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393; Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately, you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.