Optima XR646 HD. X-Ray imaging system.

FDA Device Recall #Z-1046-2025 — Class II — January 28, 2025

Recall Summary

Recall Number Z-1046-2025
Classification Class II — Moderate risk
Date Initiated January 28, 2025
Status Ongoing
Voluntary FDA Mandated

Recalling Firm

Firm GE Medical Systems, LLC
Location Waukesha, WI
Product Type Devices
Quantity 3149 (1697 US; 1452 OUS)

Product Description

Optima XR646 HD. X-Ray imaging system.

Reason for Recall

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

Distribution Pattern

Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

Lot / Code Information

GTIN: 00195278526892

Other Recalls from GE Medical Systems, LLC

Recall # Classification Product Date
Z-2154-2026 Class II Revolution Apex, X-ray Computed Tomography, Mod... Mar 26, 2026
Z-2153-2026 Class II Revolution CT ES, X-ray Computed Tomography, Mo... Mar 26, 2026
Z-2149-2026 Class II Revolution Ascend, computed tomography, Model N... Mar 26, 2026
Z-2151-2026 Class II Revolution Apex Elite, X-ray Computed Tomograph... Mar 26, 2026
Z-2150-2026 Class II Revolution CT, x-ray computed tomography, Model... Mar 26, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.