Optima XR220amx, Mobile Digital-Ready Radiographic System

FDA Device Recall #Z-2727-2026 - Class II, June 5, 2026

Recall Summary

Recall Number Z-2727-2026
Classification Class II - Moderate risk
Date Initiated June 5, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm GE Medical Systems, LLC
Location Waukesha, WI
Product Type Devices
Quantity 38 units

Product Description

Optima XR220amx, Mobile Digital-Ready Radiographic System

Reason for Recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Distribution Pattern

Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.

Lot / Code Information

UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4

Other Recalls from GE Medical Systems, LLC

Recall # Classification Product Date
Z-2728-2026 Class II Optima XR240amx, Mobile Digital Radiographic Sy... Jun 5, 2026
Z-2726-2026 Class II Optima XR200amx, Mobile Digital-Ready Radiograp... Jun 5, 2026
Z-2475-2026 Class II Giraffe OmniBed, a combination of an infant inc... May 1, 2026
Z-2477-2026 Class II Giraffe OmniBed Field Replaceable Unit (FRU), M... May 1, 2026
Z-2476-2026 Class II Giraffe OmniBed Carestations and Giraffe OmniBe... May 1, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.