NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01R...

FDA Device Recall #Z-1623-2025 - Class II, March 10, 2025

Recall Summary

Recall Number Z-1623-2025
Classification Class II - Moderate risk
Date Initiated March 10, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic Xomed, Inc.
Location Jacksonville, FL
Product Type Devices
Quantity 9,788

Product Description

NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE NIM4CPB1 NIM 4.0, REF: NIM4CPB1; PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB, REF: NIM4CPB1RF; SOFTWARE NIM4SWU143 UPGRADE V1.4.3, REF: NIM4SWU143; SOFTWARE NIM4SWU154 UPGRADE V1.5.4, REF: NIM4SWU154; SOFTWARE NIM4SWU164 UPGRADE V1.6.4, REF: NIM4SWU164

Reason for Recall

Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MD, MN, PA, OK, OH, FL, CO, GA, NC, KS, IL, WY, AK, NE, VA, MI, CA, NY, WI, UT, AZ, AR, TX, WA, AL, NJ, SD, MS, KY, TN, MO, OR, MT, MA, CT, NM, LA, WV, NH, DC, DE, IN, NV, IA, SC, ID, ME, HI, RI, VT, ND, GU, PR and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.

Lot / Code Information

All NIM Vital Nerve Monitoring Systems manufactured or installed with software version v1.6.4 or earlier. REF/UDI-DI: NIM4CM01/763000002978, 00763000395896, 00763000401597, 00763000528577; NIM4CM01RF/00763000002992; NIM4CPB1/763000002985, 00763000401603, 00763000395902, 00763000528584; NIM4CPB1RF/00763000003005; NIM4SWU143/00763000709341, 00763000869823; NIM4SWU154/00763000945398; NIM4SWU164/00763000974312

Other Recalls from Medtronic Xomed, Inc.

Recall # Classification Product Date
Z-2461-2025 Class II IPC Powerease System REF 2300000 UDI-DI: 0061... Jul 17, 2025
Z-3130-2024 Class II Medtronic ENT Ultra Round Steel Cutting Bur, Mo... Jul 5, 2024
Z-2409-2024 Class I NIM Vital Nerve Monitoring systems (all serial ... Jun 24, 2024
Z-0696-2024 Class II NIM TriVantage EMG Endotracheal Tubes, Model (R... Nov 7, 2023
Z-1224-2022 Class II NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1)... May 3, 2022

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.