MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
FDA Device Recall #Z-2778-2026 - Class II, June 16, 2026
Recall Summary
| Recall Number | Z-2778-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | June 16, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medtronic MiniMed, Inc. |
| Location | Northridge, CA |
| Product Type | Devices |
| Quantity | 3,465 |
Product Description
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
Reason for Recall
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
Distribution Pattern
Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
Lot / Code Information
Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
Other Recalls from Medtronic MiniMed, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2981-2026 | Class II | MiniMed 670G Insulin Pump, Model: MMT-1781, MMT... | Jul 13, 2026 |
| Z-2980-2026 | Class II | MiniMed 640G Insulin Pump, Model: MMT-1711, MMT... | Jul 13, 2026 |
| Z-2982-2026 | Class II | MiniMed 780G Insulin Pump, Model: MMT-1885CLW, ... | Jul 13, 2026 |
| Z-2698-2026 | Class II | MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1... | May 27, 2026 |
| Z-1747-2026 | Class II | MiniMed 620G Insulin Pump (MMT-1750) | Feb 13, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.