MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
FDA Device Recall #Z-0722-2024 — Class II — December 7, 2023
Recall Summary
| Recall Number | Z-0722-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | December 7, 2023 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Cook Medical Llc |
| Location | Bloomington, IN |
| Product Type | Devices |
| Quantity | 14 US; 101 OUS |
Product Description
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Reason for Recall
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Distribution Pattern
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
Lot / Code Information
UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316
Other Recalls from Cook Medical Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0809-2021 | Class II | Disposable Humidification Flask, Reference Part... | Dec 16, 2020 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.