Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 ...
FDA Device Recall #Z-2991-2026 - Class II, July 13, 2026
Recall Summary
| Recall Number | Z-2991-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | July 13, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medline Industries, LP |
| Location | Northfield, IL |
| Product Type | Devices |
| Quantity | 20 units |
Product Description
Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit
Reason for Recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Panama.
Lot / Code Information
All lot codes
Other Recalls from Medline Industries, LP
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2993-2026 | Class II | Medline breathing circuits labeled as: 1) ME... | Jul 13, 2026 |
| Z-2992-2026 | Class II | Medline breathing circuits labeled as: 1) HU... | Jul 13, 2026 |
| Z-3027-2026 | Class II | Medline convenience kits labeled as: 1) LAVH... | Jul 10, 2026 |
| Z-3021-2026 | Class II | Medline convenience kits labeled as: MAJOR HEA... | Jul 10, 2026 |
| Z-3020-2026 | Class II | Medline convenience kits labeled as: 1) OPEN... | Jul 10, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.