LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Euro...

FDA Device Recall #Z-2049-2026 - Class II, March 30, 2026

Recall Summary

Recall Number Z-2049-2026
Classification Class II - Moderate risk
Date Initiated March 30, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Boston Scientific Corporation
Location Saint Paul, MN
Product Type Devices
Quantity 148 units

Product Description

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5

Reason for Recall

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Belgium.

Lot / Code Information

LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284 145336 145684 146709 147185 147618 148972 149028 153330 153628 155199 155735 155898 156145 156895 157872 158926 159162 160236 160471 160615 161307 161762 162801 162821 163556 163802 167866 169673 169701 171284 172766 175538

Other Recalls from Boston Scientific Corporation

Recall # Classification Product Date
Z-2962-2026 Class II Suture Cinch Long 1pk. Endoscopic tissue app... Jul 14, 2026
Z-2961-2026 Class II OverStitch Suture Cinch, 1pk. Endoscopic tissue... Jul 14, 2026
Z-2977-2026 Class I Boston Scientific ENROUTE Transcarotid Neuropro... Jul 9, 2026
Z-2978-2026 Class I Boston Scientific ENROUTE Transcarotid Neuropro... Jul 9, 2026
Z-2838-2026 Class II Radial Jaw 4 Standard Capacity Biopsy Forceps w... Jun 30, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.