LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 646...

FDA Device Recall #Z-2048-2026 - Class II, March 30, 2026

Recall Summary

Recall Number Z-2048-2026
Classification Class II - Moderate risk
Date Initiated March 30, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Boston Scientific Corporation
Location Saint Paul, MN
Product Type Devices
Quantity 2 units

Product Description

LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5

Reason for Recall

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Belgium.

Lot / Code Information

LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787

Other Recalls from Boston Scientific Corporation

Recall # Classification Product Date
Z-2962-2026 Class II Suture Cinch Long 1pk. Endoscopic tissue app... Jul 14, 2026
Z-2961-2026 Class II OverStitch Suture Cinch, 1pk. Endoscopic tissue... Jul 14, 2026
Z-2977-2026 Class I Boston Scientific ENROUTE Transcarotid Neuropro... Jul 9, 2026
Z-2978-2026 Class I Boston Scientific ENROUTE Transcarotid Neuropro... Jul 9, 2026
Z-2838-2026 Class II Radial Jaw 4 Standard Capacity Biopsy Forceps w... Jun 30, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.