Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
FDA Device Recall #Z-1487-2020 - Class II, December 30, 2019
Recall Summary
| Recall Number | Z-1487-2020 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 30, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips North America, LLC |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 70 |
Product Description
Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Reason for Recall
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Distribution Pattern
Nationwide domestic distribution. Foreign distribution worldwide.
Lot / Code Information
Serial Codes 3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279
Other Recalls from Philips North America, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0926-2022 | Class II | The IFU for the Autoclavable Temperature Probes... | Feb 7, 2022 |
| Z-1326-2021 | Class II | Philips SureSigns VM4, VM6 and VM8 - Product Us... | Mar 23, 2021 |
| Z-0852-2021 | Class II | Emission Computed Tomography System Image Proce... | Nov 19, 2020 |
| Z-0284-2021 | Class II | Philips Sterilizable Defibrillator Paddles, Swi... | Sep 14, 2020 |
| Z-0285-2021 | Class II | Philips Sterilizable Defibrillator Paddles, Swi... | Sep 14, 2020 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.