Recall Summary
| Recall Number | Z-2549-2026 |
| Classification | Class I - Serious risk |
| Date Initiated | May 21, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Baxter Healthcare Corporation |
| Location | Deerfield, IL |
| Product Type | Devices |
Product Description
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason for Recall
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Canada.
Lot / Code Information
UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Other Recalls from Baxter Healthcare Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2551-2026 | Class I | Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Singl... | May 21, 2026 |
| Z-2548-2026 | Class I | Hillrom VOLARA System P.CIRCUIT KIT (Single Pat... | May 21, 2026 |
| Z-2550-2026 | Class I | Hillrom VOLARA System P.CIRCUIT KIT AC (Single ... | May 21, 2026 |
| Z-2262-2026 | Class II | Brand Name: Exacta Mix Product Name: Vented Mi... | Apr 22, 2026 |
| Z-2638-2026 | Class II | Adapter for Head Positioning, Model/Catalog Num... | Apr 21, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.