Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic...
FDA Device Recall #Z-0946-2015 — Class II — November 24, 2014
Recall Summary
| Recall Number | Z-0946-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 24, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Healthcare Diagnostics, Inc. |
| Location | Newark, DE |
| Product Type | Devices |
| Quantity | 1230 |
Product Description
Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the Dimension Vista System.
Reason for Recall
Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.
Distribution Pattern
Worldwide Distribution: US (nationwide) and country of: Canada.
Lot / Code Information
Lot 4FD085, exp. 12-01-2014
Other Recalls from Siemens Healthcare Diagnostics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1980-2026 | Class II | Dimension Creatinine Flex reagent cartridge. Ma... | Mar 18, 2026 |
| Z-1657-2026 | Class II | Atellica CH Enzymatic Hemoglobin A1c (A1c_E); ... | Feb 23, 2026 |
| Z-1484-2026 | Class II | Atellica CH Urine Albumin (UAlb). Material Numb... | Jan 20, 2026 |
| Z-0491-2026 | Class II | IMMULITE 2000 Oak Mix Specific Allergen; Sieme... | Oct 6, 2025 |
| Z-0492-2026 | Class II | 3gAllergy Specific IgE Universal Kit; Kit Comp... | Oct 6, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.