Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON C...

FDA Device Recall #Z-0745-2013 - Class III, December 20, 2012

Recall Summary

Recall Number Z-0745-2013
Classification Class III, Low risk
Date Initiated December 20, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics, Inc.
Location Newark, DE
Product Type Devices
Quantity 5999

Product Description

Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.

Reason for Recall

Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.

Distribution Pattern

Nationwide Distribution.

Lot / Code Information

Siemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.

Other Recalls from Siemens Healthcare Diagnostics, Inc.

Recall # Classification Product Date
Z-2975-2026 Class II Syva EMIT 2000 Quinidine Assay REF: 4Q019UL C... Jul 13, 2026
Z-2976-2026 Class II Beckman Coulter, Emit 2000 Quinidine Assay REF... Jul 13, 2026
Z-2890-2026 Class II Atellica CH Urine Albumin REF: 11537225, in vit... Jul 6, 2026
Z-2880-2026 Class III Syva Emit Disopyramide Assay Enzyme Immunoassa... Jun 26, 2026
Z-2881-2026 Class III Syva Emit tox Serum Tricyclic Antidepressant Ca... Jun 26, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.