Digital Angiography System Bransist safire Generator Model D150GC-40
FDA Device Recall #Z-0654-2022 - Class II, December 16, 2021
Recall Summary
| Recall Number | Z-0654-2022 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 16, 2021 |
| Status | Ongoing |
| Voluntary | FDA Mandated |
Recalling Firm
| Firm | Shimadzu Medical Systems |
| Location | Torrance, CA |
| Product Type | Devices |
| Quantity | 18 |
Product Description
Digital Angiography System Bransist safire Generator Model D150GC-40
Reason for Recall
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Distribution Pattern
U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.
Lot / Code Information
Generator Model: D150GC-40 Serial Number Range - 0161Q81103 to 3M99D3A55001 Specific Generator Serial Number/SMS ID: 3M99D3A24001/SMS20120038 3M99D3A4B001/SMS20140084 3M99D3A55001/SMS20150030 0161Q83701/SMS20080542 0161Q82204/SMS20080045 0161Q82302/SMS20080043 0161Q82304/SMS20080046 0261Q80901/SMS20100011 3M7A1600B001/SMS20080286 3M99D2622001/SMSMA110079 0161Q84501/SMS20090038 3M7A1600C001/SMS20200285 3M7A16A1A001/SMSMA110013 3M99D3A1C001/SMS20110202 0161Q84103/SMS20090099 0161Q83603/SMS20080505 0161Q83702/SMS20080580 0161Q83703/SMS20080581
Other Recalls from Shimadzu Medical Systems
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2294-2026 | Class II | Digital Angiography System Trinias:UD150G-40 / ... | Apr 20, 2026 |
| Z-0869-2022 | Class II | MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 | Jan 10, 2022 |
| Z-0867-2022 | Class II | MODEL: X-RAY R/F SYSTEM FLUOROspeed X1 | Jan 10, 2022 |
| Z-0868-2022 | Class II | MODEL: X-RAY TV SYSTEM SONIALVISION G4 | Jan 10, 2022 |
| Z-0655-2022 | Class II | Digital Angiography System Shimadzu Trinias ... | Dec 16, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.