CareLink Clinic, REF: MMT-7350

FDA Device Recall #Z-0203-2027 - Class II, September 11, 2026

Recall Summary

Recall Number Z-0203-2027
Classification Class II - Moderate risk
Date Initiated September 11, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic MiniMed, Inc.
Location Northridge, CA
Product Type Devices
Quantity 1

Product Description

CareLink Clinic, REF: MMT-7350

Reason for Recall

CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).

Distribution Pattern

US Nationwide distribution in the state of KS.

Lot / Code Information

UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026.

Other Recalls from Medtronic MiniMed, Inc.

Recall # Classification Product Date
Z-0265-2027 Class II Generic Name: MiniMed Go App Generic Name: Min... Sep 2, 2026
Z-0088-2027 Class II Brand Name: MiniMed Flex Product Name: MiniMed ... Aug 20, 2026
Z-2982-2026 Class II MiniMed 780G Insulin Pump, Model: MMT-1885CLW, ... Jul 13, 2026
Z-2980-2026 Class II MiniMed 640G Insulin Pump, Model: MMT-1711, MMT... Jul 13, 2026
Z-2981-2026 Class II MiniMed 670G Insulin Pump, Model: MMT-1781, MMT... Jul 13, 2026

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.