CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not f...

FDA Device Recall #Z-1062-2015 — Class II — December 16, 2014

Recall Summary

Recall Number Z-1062-2015
Classification Class II — Moderate risk
Date Initiated December 16, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maquet Datascope Corp - Cardiac Assist Division
Location Mahwah, NJ
Product Type Devices
Quantity 838 units-US; 364 units - OUS

Product Description

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Reason for Recall

Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.

Distribution Pattern

Worldwide Distribution

Lot / Code Information

Sensation 7 fr.34 cc IAB Demonstration Unit 0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469 Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470 Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units 0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568 Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576 Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272

Other Recalls from Maquet Datascope Corp - Cardiac Assis...

Recall # Classification Product Date
Z-0100-2019 Class I Cardiosave Rescue IABP, Part Number 0998-UC-080... Sep 20, 2018
Z-0099-2019 Class I Cardiosave Hybrid IABP, Part Number 0998-00-080... Sep 20, 2018
Z-1928-2018 Class I CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IA... May 3, 2018
Z-3085-2017 Class I CS 100i Intra-Aortic Balloon Pump. It suppor... Jul 17, 2017
Z-3086-2017 Class I CS 100 Intra-Aortic Balloon Pump. It support... Jul 17, 2017

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.