CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 2...

FDA Device Recall #Z-3212-2026 - Class II, August 20, 2026

Recall Summary

Recall Number Z-3212-2026
Classification Class II - Moderate risk
Date Initiated August 20, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm ICU Medical, Inc.
Location Lake Forest, IL
Product Type Devices
Quantity 185,688 units

Product Description

CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24

Reason for Recall

Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.

Lot / Code Information

21-7600-24 UDI-DI 30610586044001 Lots 6026878 6037784 6054015 6062724 6070616 6082332 6092886 6101906 6108556 6126029 6140893 6143454 6147696 6147697 6162769 6155095 6166032 6171587 6171586 21-7609-24 UDI-DI 30610586044018 Lot 6022079 6037697 6037698 6053978 6061525 6072970 6072964 6070575 6077773 6092846 6101798 6108514 6108515 6108516 6140862 6140861 6140865 6155132

Other Recalls from ICU Medical, Inc.

Recall # Classification Product Date
Z-2724-2026 Class II ICU Medical Plum Solo Precision IV Pump, v1.0.3... Jun 10, 2026
Z-2723-2026 Class II ICU Medical Plum Duo Infusion Pump, v1.2.3: 1)... Jun 10, 2026
Z-3163-2026 Class II IV Sets with drip chambers May 12, 2026
Z-3162-2026 Class II IV Sets with drip chambers May 12, 2026
Z-3157-2026 Class II IV Non-Dedicated Gravity Administration Sets wi... May 12, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.