| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Nov 30, 2023 | GammaPod - Treatment Planning System-a component of the GammaPod Irradiation ... | Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS gen... | Class II | Xcision Medical Systems, LLC |
| Jul 3, 2019 | GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Pr... | One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment ... | Class II | Xcision Medical Systems, LLC |
Frequently Asked Questions
When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.
All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.
Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.