Browse Device Recalls

3 FDA medical device safety recalls.

FDA Medical Device Recall Enforcement Database

Browse 3 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.

Medical Device Recalls

Browse 3 FDA device recalls.

Clear
DateProductReasonClassFirm
May 9, 2018 OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.... Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specificall... Class II Johnson & Johnson Surgical Vision Inc
Feb 20, 2015 Catalys Precision Laser System-Catalys - U - US product; Catalys-1 - Internat... Software anomaly on OptiMedica CATALYS System version 3.00.05 that may result in incorrect parame... Class II Optimedica Corporation
Dec 20, 2013 Catalys Precision Laser System (United States) - Catalys-U: Catalys Precisio... Potential for unexpected heating of the transformer mounting bolt located within the system chass... Class II Optimedica Corporation

Frequently Asked Questions

When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.

All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.

Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.