| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| May 6, 2013 | da Vinci 5 mm Instrument Cannulae for use with the da Vinci S and Si Surgical... | Intuitive Surgical is implementing a voluntary Field Action to implement a modified inspection pr... | Class II | Intuitive Surgical, Inc. |
| May 2, 2013 | Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417)... | The product has the potential for false negative or delayed reaction for Arginine (ARG) results. ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 1, 2013 | Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers, Model Numbers... | Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix¿ and Direct ... | Class II | Applied Medical Resources Corp |
| Apr 30, 2013 | TX1 Tissue Removal System disposable handpieces. Product Usage: The TX1 T... | American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable han... | Class II | American Optisurgical Inc |
| Apr 30, 2013 | uTASWako DCP, Lot TG340 2013.11, REF 995-60701, Wako Pure Chemical Industries... | Product fails to meet the accuracy at the low end of the measuring range. | Class II | Wako Life Sciences, Inc. |
| Apr 30, 2013 | FAST Procedure Kit. Intended for use as a Minor Surgical Procedures Tray. | American Optisurgical, Inc initiated a recall of ChloraPrep Hi-Lite Orange 3mL Applicator, in re... | Class II | American Optisurgical Inc |
| Apr 26, 2013 | AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Per... | Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it ... | Class II | Beckman Coulter Inc. |
| Apr 22, 2013 | STERRAD CYCLESURE 24 Biological Indicator (BI) The STERRAD CYCLESURE 24 Bi... | Advanced Sterilization Products is recalling certain lots of STERRAD CYCLESURE 24 Biological Indi... | Class II | Advanced Sterilization Products |
| Apr 18, 2013 | Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI ... | Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical micros... | Class II | Carl Zeiss Meditec, Inc. |
| Apr 17, 2013 | Synchron LX System(s) and UniCel DxC 600/800 System(s) Lactate Reagent, Part ... | Beckman Coulter initiated the recall for the Lactate Assay (LACT) Lot M206209 as they received a ... | Class II | Beckman Coulter Inc. |
| Apr 17, 2013 | Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Nu... | A complaint investigation has found that product fractures can occur at the distal end of the cat... | Class II | Xlumena, Inc. |
| Apr 16, 2013 | Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereT... | The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG11101... | Class II | Alere San Diego, Inc. |
| Apr 15, 2013 | Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Cont... | The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which ori... | Class III | Bio-rad Laboratories, Inc. |
| Apr 15, 2013 | FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic ... | AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Gluco... | Class I | Abbott Diabetes Care, Inc. |
| Apr 5, 2013 | Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batterie... | Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instrumen... | Class I | Newport Medical Instruments Inc |
| Apr 4, 2013 | QuietCare-Networked. Intended for use in monitoring the environmental cond... | Customers at 4 facilities reported receiving duplicate alerts using the QuietCare-Networked product. | Class II | Intel-GE Care Innovations LLC |
| Apr 4, 2013 | QuietCare. Intended for use in monitoring the environmental conditions and... | Customers at 4 facilities reported receiving duplicate alerts using the QuietCare-Networked product. | Class II | Intel-GE Care Innovations LLC |
| Apr 4, 2013 | Software correction applies to customers who use the RT Therapist RTT4.2.108... | Software update to fix multiple safety related issues. | Class II | Siemens Medical Solutions USA, Inc |
| Apr 3, 2013 | TOSHIBA Kalare Diagnostic X-Ray System K110785 | Firm initiated recall because when radiographic conditions with the small focus are set and the r... | Class II | Toshiba American Medical Systems Inc |
| Apr 2, 2013 | Product is distributed by Zimmer (Warsaw, Indiana)/Synvasive Technology, Inc.... | Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors in... | Class II | Synvasive Technology Inc |
| Mar 30, 2013 | American Optisurgical's TX1 Tissue Removal System Console. The console is a s... | The recall has been initiated due to an overheating protection voltage suppressor on the redundan... | Class II | American Optisurgical Inc |
| Mar 28, 2013 | Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that ... | When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be ... | Class II | Siemens Medical Solutions USA, Inc. |
| Mar 28, 2013 | EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to a... | Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor ... | Class II | Advanced Sterilization Products |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finis... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 1 Hole, Metal F... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip with No Holes, Non-R... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Vented, Metal Finis... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 2mm X 34cm Regular Tip with No Hole, Metal ... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 10mm X 32cm Regular Tip, vented, Non-Reflec... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finis... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 28cm Regular Tip, vented, Non-Reflect... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip , Non-Vented, Metal ... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Aspiration Needle Tip. Suctio... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, Vented, Metal Finis... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, vented, Non-Reflect... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Tip Hydrodissection Tip, Metal F... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: Vented Tip, 5mm X 45cm Non-Reflective with ... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflect... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal F... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Non-Vented, Metal F... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 10mm X 32cm Pool Suction Tip, Metal Finish.... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 32cm Hydrodissection Tip, Non-Reflect... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finis... | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incor... | Class II | Stryker Endoscopy |
| Mar 27, 2013 | RoboCouch Patient Positioning System Product Usage: is intended for use... | A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) moti... | Class II | Accuray Incorporated |
| Mar 26, 2013 | Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option,... | Varian has identified an anomaly with the RPM Respiratory Gating System version I.7.5 with 3D Op... | Class II | Varian Medical Systems, Inc. Oncology Systems |
| Mar 22, 2013 | STERRAD 100NX Sterilization System, P/N 10104 Product Usage: The STERRAD ... | The recall was initiated because Advanced Sterilization Products (ASP) has determined that the ST... | Class II | Advanced Sterilization Products |
| Mar 22, 2013 | EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to auto... | Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufa... | Class II | Advanced Sterilization Products |
| Mar 22, 2013 | Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers: MMT-511, ... | Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap m... | Class II | Medtronic MiniMed |
| Mar 22, 2013 | STERRAD 200 Sterilization System, P/N 10201 Product Usage: The STERRAD 20... | The recall was initiated because Advanced Sterilization Products (ASP) has determined that the ST... | Class II | Advanced Sterilization Products |
| Mar 20, 2013 | AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Che... | Beckman Coulter, Inc. initiated this recall because there is a software error that has the potent... | Class II | Beckman Coulter Inc. |