| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Feb 3, 2025 | HemosIL LMW Heparin Controls; Part Number: 0020300200; | Multiple complaints indicating lower than expected quality control (QC) results and in some insta... | Class II | Instrumentation Laboratory |
| Feb 3, 2025 | Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400... | Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared... | Class II | Beckman Coulter, Inc. |
| Feb 3, 2025 | Brand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS i... | Potential variability in calculations from fast array scans compared to array scans when operatin... | Class II | Medimaps Group |
| Feb 3, 2025 | CardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital S... | Due to issues associated with data migration of patient information and results to a Cloud servic... | Class II | St. Jude Medical |
| Jan 31, 2025 | Numerous models of nonsterile hemostatic forceps: (1) REF BH198R, Fuchsig ... | The forceps have been used in ways not covered by the design resulting in breakage of the clamps. | Class II | Aesculap Inc |
| Jan 31, 2025 | MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MM... | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Optima MR360, NMRI system | For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute a... | Class II | GE Healthcare (China) Co., Ltd. |
| Jan 31, 2025 | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tu... | Due to the inner diameter of the suction tubing connector being smaller than the suction canister... | Class II | DeRoyal Industries Inc |
| Jan 31, 2025 | Brand Name: Banner Health Systems Product Name: Lap Pack - 170307 Model/Cat... | Sterility assurance with procedure trays | Class II | American Contract Systems, Inc. |
| Jan 31, 2025 | SIGNA MR380, NMRI system, Non-US Only | For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute a... | Class II | GE Healthcare (China) Co., Ltd. |
| Jan 31, 2025 | Numerous models of nonsterile hemostatic forceps: (1) REF FB458R, Glover A... | The forceps have been used in ways not covered by the design resulting in breakage of the clamps. | Class II | Aesculap Inc |
| Jan 31, 2025 | MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851) | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 740G (MMT-1811, MMT-1812, MMT-1861, MMT-1862) | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MM... | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT... | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 640G (MMT-1711, MMT-1712, MMT-1751, MMT-1752) | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Paradigm insulin pump, REF: MMT-754 | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 620G (MMT-1750) | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Numerous models of nonsterile hemostatic forceps: (1) REF BH100R, Carrel Art... | The forceps have been used in ways not covered by the design resulting in breakage of the clamps. | Class II | Aesculap Inc |
| Jan 31, 2025 | Paradigm REF: MMT-712 and MMT-715 | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 720G (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MM... | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Numerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf ... | The forceps have been used in ways not covered by the design resulting in breakage of the clamps. | Class II | Aesculap Inc |
| Jan 31, 2025 | Brand Name: Banner Health Systems Product Name: Cysto Pack Model/Catalog Nu... | Sterility assurance with procedure trays | Class II | American Contract Systems, Inc. |
| Jan 31, 2025 | SIGNA MR355, SIGNA MR360, NMRI system | For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute a... | Class II | GE Healthcare (China) Co., Ltd. |
| Jan 31, 2025 | MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755) | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | SIGNA Creator, SIGNA Explorer, NMRI systems | For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute a... | Class II | GE Healthcare (China) Co., Ltd. |
| Jan 31, 2025 | Numerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus To... | The forceps have been used in ways not covered by the design resulting in breakage of the clamps. | Class II | Aesculap Inc |
| Jan 31, 2025 | MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MM... | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were foun... | Class II | Medtronic MiniMed, Inc. |
| Jan 31, 2025 | Brivo MR355, NMRI system | For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute a... | Class II | GE Healthcare (China) Co., Ltd. |
| Jan 31, 2025 | Brand Name: Banner University Med Ctr. Product Name: Cardiac Cath Lab Pack... | Sterility assurance with procedure trays | Class II | American Contract Systems, Inc. |
| Jan 30, 2025 | Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009; | Telcare, LLC has received customer complaints related to battery expansion/swelling for its Phili... | Class II | Telcare, LLC |
| Jan 30, 2025 | Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: P... | Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wal... | Class I | Micro Therapeutics, Inc. |
| Jan 30, 2025 | Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog N... | Firm became aware of software issues related to most recent firmware update which could result in... | Class II | LEICA BIOSYSTEMS NUSSLOCH GMBH |
| Jan 30, 2025 | Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: P... | Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wal... | Class I | Micro Therapeutics, Inc. |
| Jan 29, 2025 | PENTAX Medical Video Colonoscope- Intended to provide optical visualization ... | During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c s... | Class II | Pentax of America Inc |
| Jan 29, 2025 | PENTAX Medical Video Upper GI Scope-Intended to provide optical visualization... | During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c s... | Class II | Pentax of America Inc |
| Jan 29, 2025 | Pentax Medical Video Processor- Intended to be used with PENTAX Medical endos... | During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c s... | Class II | Pentax of America Inc |
| Jan 29, 2025 | Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643... | Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire rema... | Class II | Stryker Corporation |
| Jan 28, 2025 | Discovery XR656HD. X-Ray imaging system. | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AE... | Class II | GE Medical Systems, LLC |
| Jan 28, 2025 | Discovery XR656 HD. X-Ray imaging system. | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AE... | Class II | GE Medical Systems, LLC |
| Jan 28, 2025 | Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00... | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AE... | Class II | GE Medical Systems, LLC |
| Jan 28, 2025 | Optima XR646 HD. X-Ray imaging system. | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AE... | Class II | GE Medical Systems, LLC |
| Jan 27, 2025 | Osstem Implant System - Abutment: ET Multi AbutmentSize: 4.8D 5.0GH - Abutme... | Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments ... | Class II | Hiossen Inc. |
| Jan 27, 2025 | TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP | Tissue approximation straps (zip-ties) have broken before or during implantation so firm develope... | Class II | TAS Medical Inc |
| Jan 27, 2025 | Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001 | Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucos... | Class II | Dexcom, Inc. |
| Jan 27, 2025 | ET Rigid Abutment (Mini) Size: 4.5D 2.0G/H 4.0H- Intended for use with a dent... | Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments ... | Class II | Hiossen Inc. |
| Jan 27, 2025 | ET Rigid Abutment (Mini) SIze: 4.5D 3.0G/H 7.0H - Intended for use with a den... | Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments ... | Class II | Hiossen Inc. |
| Jan 27, 2025 | ET Rigid Abutment (Mini) Size 4.5D 1.0G/H 5.0H - Intended for use with a den... | Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments ... | Class II | Hiossen Inc. |
| Jan 27, 2025 | Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001... | Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucos... | Class II | Dexcom, Inc. |
| Jan 27, 2025 | Prelude IDEAL 4 F Hydrophilic Sheath Introducer, REF: PID4F16021PW/D | Hydrophilic Sheath Introducer labeled for 4F dilators may instead incorrectly contain 5F dilators... | Class II | Merit Medical Systems, Inc. |