| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| May 23, 2016 | PrismaSate BGK2/0 Dialysis Solution for Continuous Renal Replacement Therapy | Presence of leaks near top of the PrismaSate bags | Class II | Baxter Healthcare Corp. |
| May 19, 2016 | Conical portio adapter (acorn) small & large are used for examination and dia... | The distal end of the Conical portio adapter (acorn) is composed of a small cap (head) that parti... | Class II | Richard Wolf Medical Instruments Corp. |
| May 11, 2016 | Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of... | There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup... | Class II | Medline Industries Inc |
| May 6, 2016 | Ariol is an automated scanning microscope and image analysis system. It is in... | An issue during the manufacturing process caused systems to have an improperly activated Windows ... | Class III | Leica Biosystems Richmond Inc. |
| May 2, 2016 | CytoVision Image Analysis and Capture System, an automated cell-locating devi... | Systems have an improperly activated Windows 7 OS, even though a valid Windows 7 license was asso... | Class III | Leica Biosystems Richmond Inc. |
| Apr 21, 2016 | GEM(TM), FLOWCOUPLER(R), 4.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2... | Instructions for use booklet may puncture the outer Tyvek lid. | Class II | Baxter Healthcare Corp. |
| Apr 21, 2016 | GEM(TM), FLOWCOUPLER(R), 3.5 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2... | Instructions for use booklet may puncture the outer Tyvek lid. | Class II | Baxter Healthcare Corp. |
| Apr 21, 2016 | GEM(TM), FLOWCOUPLER(R), 3.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2... | Instructions for use booklet may puncture the outer Tyvek lid. | Class II | Baxter Healthcare Corp. |
| Apr 21, 2016 | GEM(TM), FLOWCOUPLER(R), 2.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2... | Instructions for use booklet may puncture the outer Tyvek lid. | Class II | Baxter Healthcare Corp. |
| Apr 21, 2016 | GEM(TM), FLOWCOUPLER(R), 2.5 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2... | Instructions for use booklet may puncture the outer Tyvek lid. | Class II | Baxter Healthcare Corp. |
| Apr 5, 2016 | Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) ... | Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up ... | Class II | Leica Microsystems, Inc. |
| Apr 5, 2016 | Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the... | Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the e... | Class II | Leica Microsystems, Inc. |
| Mar 31, 2016 | E.cam or Symbia systems that use foresight detectors - Product Usage: To dete... | E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may los... | Class II | Siemens Medical Solutions USA, Inc. |
| Mar 25, 2016 | The Plum 360 is a large volume infuser capable of delivering fluids for a var... | Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were att... | Class II | Hospira Inc. |
| Mar 25, 2016 | The Plum A+ is a dual-line volumetric infusion system designed to meet the gr... | Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were att... | Class II | Hospira Inc. |
| Mar 24, 2016 | The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic ... | Interaction of Fulvestrant with the ARCHITECT Estradiol assay, leading to falsely elevated estrad... | Class II | Abbott Laboratories |
| Mar 22, 2016 | Novocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566. ... | Novocastra Liquid Mouse Monoclonal Antibody Calretinin when used in combination with specific Lot... | Class II | Leica Microsystems, Inc. |
| Feb 29, 2016 | Gomco Style Circumcision Clamp Trays is an instrument used to compress the fo... | Two lots of Gomco Style Circumcision Clamp Trays, SSI-0034 and 10-1100 contained a 1.3cm circumci... | Class II | Cardinal Health |
| Feb 29, 2016 | The Vesa 75/100 adaptation with rotation is an accessory to a surgical camera... | Supplier of the "VESA 75 / 100 adaptation with rotation", which is part of their monitor mounts a... | Class II | Richard Wolf Medical Instruments Corp. |
| Feb 26, 2016 | The E.CAM gamma camera system Product Usage: The E.CAM gamma camera syst... | We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-000... | Class II | Siemens Medical Solutions USA, Inc. |
| Feb 26, 2016 | The Symbia gamma camera system Product Usage: The Symbia gamma camera sy... | We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-000... | Class II | Siemens Medical Solutions USA, Inc. |
| Feb 18, 2016 | The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS), is an Ov... | The instructional insert for the In Touch Control Solution for use with the In Touch meter and I... | Class II | Livongo Health |
| Feb 16, 2016 | Automated peritoneal dialysis (APD) cycler | Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructi... | Class II | Baxter Healthcare Corp. |
| Jan 11, 2016 | Leica Microsystems (Schweiz) Ag. The common name of the device is surgical mi... | An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W i... | Class II | Leica Microsystems, Inc. |
| Dec 31, 2015 | The e.cam Dual Signature Gamma Camera system. Used to detect or image the di... | To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not ... | Class II | Siemens Medical Solutions USA, Inc. |
| Dec 30, 2015 | This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody ... | The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequat... | Class II | Leica Microsystems, Inc. |
| Dec 22, 2015 | The Symbiq infusion pump is a device used in a health care facility to pump f... | Potential for the Symbiq pole clamp assemblies to be missing the rubber stop pad. | Class II | Hospira Inc. |
| Dec 21, 2015 | The Concerto Basic Shower Trolley is intended for assisted hygiene care, espe... | Reports of the bolts connecting the stretcher to the hydraulic piston and the stretcher tilting m... | Class II | Arjo, Inc. dba ArjoHuntleigh |
| Dec 3, 2015 | VS300 radiographic Cassette Holder or Wall Stand. For use in radiology depart... | The wall stand cassette holder can suddenly fall to the bottom of its travel, near the ground and... | Class II | Del Medical, Inc. |
| Dec 3, 2015 | VS200 radiographic Cassette Holder or Wall Stand. For use in radiology depart... | The wall stand cassette holder can suddenly fall to the bottom of its travel, near the ground and... | Class II | Del Medical, Inc. |
| Nov 30, 2015 | Baxter Y-Type Spike Adapter, 7" (18 cm), 1C8433, Product Usage: C8433: F... | Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike ad... | Class II | Baxter Healthcare Corp. |
| Nov 30, 2015 | Baxter CLEARLINK System, Blood Bag Spike Adapter, Luer Activated Valve for I.... | Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike ad... | Class II | Baxter Healthcare Corp. |
| Nov 30, 2015 | Baxter INTERLINK System, Y-Type Solution Set, 78 (2.0 m), 2 Injection Sites, ... | Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike ad... | Class II | Baxter Healthcare Corp. |
| Nov 27, 2015 | Bond Polymer Refine Red Detection DS9390 kits. The common name of the device ... | The Firm informed the importer/distributor, Leica Microsystems, a manufacturing error occurred d... | Class II | Leica Microsystems, Inc. |
| Nov 25, 2015 | Reuseable rotary blade/abrader, product number 899751514. The oval burr fron... | Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the dist... | Class II | Richard Wolf Medical Instruments Corp. |
| Nov 25, 2015 | Reuseable rotary blade/abrader, product number 899751504. The oval burr late... | Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the dist... | Class II | Richard Wolf Medical Instruments Corp. |
| Nov 25, 2015 | Reuseable rotary blade/abrader, product number 899751305. The round burr is... | Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the dist... | Class II | Richard Wolf Medical Instruments Corp. |
| Nov 20, 2015 | WARM STEAM Personal Steam Inhaler-Adjustable control to regulate flow of stea... | "The public reason for the recall is that the product is being recalled out of an abundance of ca... | Class II | Bigwall Enterprises |
| Oct 6, 2015 | Baxter 15L Cycler Drainage Bag Sterile, nonpyrogenic fluid path | Leakage from the inlet port of the 15L Cycler Drainage Bags. | Class II | Baxter Healthcare Corp. |
| Sep 28, 2015 | The Centricity PACS-IW with Universal Viewer is a device that displays medica... | Inaccurate distance measurements with magnified projection X-ray images. | Class II | GE Healthcare |
| Sep 28, 2015 | The Centricity Universal Viewer Version is a device that displays medical ima... | Inaccurate distance measurements with magnified projection X-ray images. | Class II | GE Healthcare |
| Sep 16, 2015 | Centricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device tha... | Images may be missing when a system parameter MapRoute is set to a value greater than 1. | Class II | GE Healthcare |
| Sep 3, 2015 | Novocastra Lyophilized Mouse Monoclonal Antibody CD10 (CD10). Immunohistochem... | These Lot numbers are not stable up to the expiry date on the product labeling. | Class II | Leica Microsystems, Inc. |
| Aug 25, 2015 | Ambitex NMD400 Nitrile Exam Gloves size medium. General Hospital and Personal... | Ambitex Nitrile NMD400 Exam Gloves size medium on exam hold were inadvertently placed into commer... | Class II | Cardinal Health |
| Aug 11, 2015 | Centricity PACS-IW with Universal Viewer Product Usage: The Centricity PA... | Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be ... | Class II | GE Healthcare |
| Aug 11, 2015 | Centricity Universal Viewer Product Usage: The Centricity Universal Viewe... | Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be ... | Class II | GE Healthcare |
| Aug 5, 2015 | The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departme... | The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm ... | Class III | Leica Microsystems, Inc. |
| Aug 3, 2015 | Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Wee... | The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying ... | Class II | Medline Industries Inc |
| Jul 31, 2015 | Siemens Symbia T Series, a radiological imaging system that is a combination ... | Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbi... | Class II | Siemens Medical Solutions USA, Inc. |
| Jul 31, 2015 | Siemens Intevo Series system, a radiological imaging system that is a combina... | Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbi... | Class II | Siemens Medical Solutions USA, Inc. |