| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Nov 26, 2012 | NEONATAL TOTAL GALACTOSE KIT Product Numb... | Truncated V-Bottomed Microplate Wells used in assays contain defective wells. | Class II | Perkinelmer |
| Nov 21, 2012 | Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemis... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | 18 Fr 6mm Sft Flow Angled Wire The Sarns Aortic Cannula is indicated for u... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | 24 Fr 8mm Soft Flow Ang Wire W/L The Sarns Aortic Cannula is indicated for... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | 18 Fr 6mm Sft Flow Ang Wire W/L The Sarns Aortic Cannula is indicated for ... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | Oxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mc... | Boxes labeled as CIP5 may contain cartridges of AMP5. | Class III | Remel Inc |
| Nov 21, 2012 | Siemens Chek-Stix Combo Pak Catalog Number: 1364 Chek-Stix¿ Positive and... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | Custom Procedural Trays: Merit's Custom Angio Kits A custom procedural tra... | Custom Procedural trays contain Medrad Hand Controller Sheaths which were recalled because the se... | Class II | Merit Medical Systems, Inc. |
| Nov 21, 2012 | Siemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364N ... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class II | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | 24 Fr 8mm Sft Flow Straight The Sarns Aortic Cannula is indicated for use ... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry (... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | Siemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364M ... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry C... | When these lots of control material are reconstituted, the positive control solution leukocyte re... | Class III | Siemens Healthcare Diagnostics Inc |
| Nov 21, 2012 | 24 Fr 8mm Sft Flow Str Wire W/L The Sarns Aortic Cannula is indicated for ... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | 24 Fr 8mm Sft Flow Straight Wir The Sarns Aortic Cannula is indicated for ... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 21, 2012 | 24 Fr 8mm Soft Flow Ang W/L The Sarns Aortic Cannula is indicated for use ... | The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic... | Class II | Terumo Cardiovascular Systems Corporation |
| Nov 20, 2012 | WANG Transbronchial Aspiration Needles Cat. No. MW-222, WANG Cytology Needle... | ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration... | Class II | ConMed Corporation |
| Nov 20, 2012 | The Centricity Laboratory System is intended to be an information system desi... | If environment variables MICRO_CORRECT_MSG_TYPE and CORRECT_MSG_TYPE are set to different value... | Class II | GE Healthcare It |
| Nov 20, 2012 | WANG Transbronchial Aspiration Needles Cat. No. SW-121, WANG Cytology Needle... | ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration... | Class II | ConMed Corporation |
| Nov 20, 2012 | WANG Transbronchial Aspiration Needles Cat. No. MW-322, WANG Cytology Needle... | ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration... | Class II | ConMed Corporation |
| Nov 20, 2012 | The Centricity Laboratory System is intended to be an information system desi... | GE Healthcare is aware of a potential safety issue associated with the use of GE Centricity Labor... | Class III | GE Healthcare It |
| Nov 20, 2012 | WANG Transbronchial Aspiration Needles Cat. No. MW-221, WANG Cytology Needle... | ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration... | Class II | ConMed Corporation |
| Nov 20, 2012 | Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System... | The Centricity Laboratory removes free-text notes on a master panel when the user deletes the sla... | Class II | GE Healthcare It |
| Nov 19, 2012 | Medtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioA... | Medtronic is initiating an Urgent Medical Device Customer Notification. We have confirmed that c... | Class II | Medtronic Inc. Cardiac Rhythm Disease Management |
| Nov 19, 2012 | Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing ... | FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes... | Class I | Fisher & Paykel Healthcare, Ltd. |
| Nov 19, 2012 | Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic ana... | A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener ... | Class III | Roche Diagnostics Operations, Inc. |
| Nov 19, 2012 | Patient Circuit used with the Life Pulse High Frequency Ventilator. The Pati... | Customer complaints received indicate the heater wire insulation can melt, causing a short which ... | Class I | Bunnell, Inc. |
| Nov 19, 2012 | Titanium Tack Starter Kit, REF 400-270, Rx Only, non-sterile, BIOHORIZONS, BI... | The cleaning instructions provided may not be sufficient to remove residual contaminants after su... | Class II | BioHorizons Implant Systems Inc |
| Nov 19, 2012 | Coonrad/Morrey New Ulnar Revision Kit for Use with Non-Interchangeable Regula... | Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inn... | Class II | Zimmer, Inc. |
| Nov 19, 2012 | AutoTac Delivery Handle, REF 400-200, Rx Only, non-sterile, BIOHORIZONS, BIRM... | The cleaning instructions provided may not be sufficient to remove residual contaminants after su... | Class II | BioHorizons Implant Systems Inc |
| Nov 19, 2012 | COONMAD/MORREY TOTAL ELBOW ELBOW COMPONENT REPLACEMENT SET BUSHINGS & PINS ... | Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inn... | Class II | Zimmer, Inc. |
| Nov 16, 2012 | Enterprise 9000 Bed; an AC-powered adjustable hospital bed with built-in elec... | ArjoHuntleigh has received reports of unintended movement where the backrest and/or thigh section... | Class II | Arjo, Inc. dba ArjoHuntleigh |
| Nov 16, 2012 | Enterprise 8000 Bed; an AC-powered adjustable hospital bed; ArjoHuntleigh AB,... | ArjoHuntleigh has received reports of unintended movement where the backrest and/or thigh section... | Class II | Arjo, Inc. dba ArjoHuntleigh |
| Nov 16, 2012 | Centricity Enterprise Archive is a software product for receiving, archiving ... | There is a potential for data loss associated with empty path names resulting from HL-7 updates w... | Class II | Ge Healthcare It |
| Nov 16, 2012 | DePuy Mitek 11 MM Fully Fluted Reamer, Sterile Product Number: 232423 DeP... | Specific units of the DePuy Mitek Sterile Reamers are not labeled correctly, 7 units were labele... | Class II | DePuy Mitek, Inc., a Johnson & Johnson Co. |
| Nov 16, 2012 | Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: ... | Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwis... | Class I | Praxair Inc. |
| Nov 16, 2012 | Clinical Chemistry Phenobarbital, list number 1E08 | Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773... | Class II | Abbott Laboratories, Inc |
| Nov 15, 2012 | Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2%... | Some disposable cannulas (with 4-digit lot numbers) provided with the Amvisc or Amvisc Plus OVD a... | Class I | Bausch and Lomb, Incorporated |
| Nov 15, 2012 | Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) an... | Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (... | Class II | Viterion TeleHealthcare Llc |
| Nov 15, 2012 | Viterion TeleHealth Monitor Model Number V100, Product Codes 09377512 (US) an... | Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (... | Class II | Viterion TeleHealthcare Llc |
| Nov 15, 2012 | Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Produ... | Biomet investigation identified that the square end of the offset tibial tray adaptor locking ins... | Class II | Biomet, Inc. |
| Nov 15, 2012 | EEG NeuroAmp, Model Number: CS 10090; CS 10137. Biofeedback and Relaxation | EEG Info, Inc. is recalling the EEG Neuroamp device because they have identified a potential risk... | Class II | EEG Info |
| Nov 15, 2012 | Viterion TeleHealth Monitor Model Number V100-BGM, Product Code 81842478 (US... | Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (... | Class II | Viterion TeleHealthcare Llc |
| Nov 15, 2012 | GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Ar... | While merging exams in a test instance, merges carry forward in the production system, when the m... | Class II | GE Healthcare It |
| Nov 15, 2012 | ConMed Linvatec Biomaterials, PopLok bone Punch PKL-35M, 3.5mm Orthope... | ConMed Linvatech recalled one lot of their PopLok bone Punch 3.5mm because the metal shaft diamet... | Class II | Linvatec Corp. dba ConMed Linvatec |
| Nov 14, 2012 | Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Ste... | All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, ... | Class II | Zimmer, Inc. |
| Nov 14, 2012 | Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonst... | All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, ... | Class II | Zimmer, Inc. |
| Nov 14, 2012 | Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 1... | Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room... | Class II | Siemens Healthcare Diagnostics, Inc |
| Nov 14, 2012 | Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Ste... | All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, ... | Class II | Zimmer, Inc. |