GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communic...
FDA Device Recall #Z-2579-2016 — Class II — November 15, 2012
Recall Summary
| Recall Number | Z-2579-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 15, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare It |
| Location | Barrington, IL |
| Product Type | Devices |
| Quantity | 498 units |
Product Description
GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.
Reason for Recall
While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.
Distribution Pattern
Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.
Lot / Code Information
Centricity PACS versions 3.1.1.x through 3.2.1.x
Other Recalls from GE Healthcare It
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0748-2014 | Class II | Merge Mammo is a multi-modality, vendor-neutral... | Aug 26, 2013 |
| Z-0116-2014 | Class II | The Centricity PACS Workstation is intended for... | Jun 11, 2013 |
| Z-0833-2015 | Class II | Centricity Enterprise Archive is a software pro... | Apr 19, 2013 |
| Z-0620-2014 | Class II | The Centricity Laboratory System is intended to... | Mar 12, 2013 |
| Z-0330-2014 | Class II | The Centricity Laboratory System is intended to... | Mar 12, 2013 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.