| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Aug 11, 2016 | RAPIDLab 1245 Blood Gas Analyzer Siemens Material Number (SMN): 10321844, 1... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 11, 2016 | Siemens RAPIDLab 1240 Blood Gas Analyzer Siemens Material Number (SMN): 103... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 11, 2016 | Siemens RAPIDPoint 500 Blood Gas Analyzer Siemens Material Number (SMN): ... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 11, 2016 | RAPIDLab 1260 Blood Gas Analyzer Siemens Material Number (SMN): 10321846, 1... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 11, 2016 | RAPIDLab 1265 Blood Gas Analyzer Siemens Material Number (SMN): 10321852, 1... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 11, 2016 | Siemens RAPIDPoint 405 Blood Gas Analyzer nBili Siemens Material Number (SMN... | There is a potential for the first and/or last name of one patient to be printed with Patient ID ... | Class II | Siemens Healthcare Diagnostics Inc |
| Aug 9, 2016 | ADVIA Centaur¿ PSA Assay; Cat No. 06574155/SMN 10310292 (100 test)/02676506/S... | Communication provided to emphasize that the PSA values should be interpreted in accordance with ... | Class II | Siemens Healthcare Diagnostics, Inc |
| Aug 8, 2016 | Siemens ADVIA Centaur¿ Calibrator U, For in vitro diagnostic use in calibrat... | Failure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from... | Class II | Siemens Healthcare Diagnostics, Inc |
| Aug 3, 2016 | 1.5 mm HYDRO LeMaitre Valvulotome, 40 cm. Sterile REF # 1010-00 Product U... | Wire which deploys the loop may be come detached at the handle causing the device to be inoperable | Class II | LeMaitre Vascular, Inc. |
| Aug 3, 2016 | 1.5 mm HYDRO LeMaitre Valvulotome, 98 cm, Sterile REF # 1009-00 Product ... | Wire which deploys the loop may be come detached at the handle causing the device to be inoperable | Class II | LeMaitre Vascular, Inc. |
| Aug 3, 2016 | 1.5 mm HYDRO LeMaitre Valvulotome without LeMills, 98 cm. Sterile REF# 1009-... | Wire which deploys the loop may be come detached at the handle causing the device to be inoperable | Class II | LeMaitre Vascular, Inc. |
| Aug 2, 2016 | SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. I... | Tip Breakage at the glue joint between the obturator body and tip | Class II | Hologic, Inc |
| Aug 2, 2016 | SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. I... | Tip Breakage at the glue joint between the obturator body and tip | Class II | Hologic, Inc |
| Jul 29, 2016 | Zyno Medical Administration Set, Sterile. Rx For Use with Zyno Medical Infus... | Administration Set potential filter leaking | Class II | Zyno Medical LLC |
| Jul 29, 2016 | Zyno Medical Administration Set, Sterile. RX For Use with Zyno Medical Infus... | Administration Set potential filter leaking | Class II | Zyno Medical LLC |
| Jul 29, 2016 | Zyno Medical Administration Set Sterile. Rx For Use with Zyno Medical Infusi... | Administration Set potential filter leaking | Class II | Zyno Medical LLC |
| Jul 29, 2016 | Zyno Medical Administration Set, Sterile. Rx For Use with Zyno Medical Infus... | Administration Set potential filter leaking | Class II | Zyno Medical LLC |
| Jul 26, 2016 | IntelliVue Patient Monitor Product Usage: The monitors are indicated for ... | If an affected Patient Monitor has been powered on continuously for several months, any displayed... | Class II | Philips Electronics North America Corporation |
| Jul 25, 2016 | Terumo Custom Cardiovascular System Cardiovascular Procedure Kits labeled con... | Terumo Custom Cardiovascular Procedure Kits are labeled as Xcoating surface coated tubing, but tu... | Class II | Terumo Cardiovascular Systems Corp |
| Jul 19, 2016 | MultiDiagnost Eleva with Flat Detector; x-ray system. Product Codes: 708034, ... | The two bolts that attach the clamping plate to the lateral shaft may break, which could cause th... | Class II | Philips Electronics North America Corporation |
| Jul 19, 2016 | Smith & Nephew Whipknot Soft Tissue Cinch (USP#5) suture. PN: 7211015 Used ... | Sterility compromised due to breach in sterile packaging | Class II | Smith & Nephew, Inc. |
| Jul 19, 2016 | UroDiagnost; x-ray system. Product Code: 708033 | The two bolts that attach the clamping plate to the lateral shaft may break, which could cause th... | Class II | Philips Electronics North America Corporation |
| Jul 19, 2016 | MultiDiagnost Eleva; x-ray system. Product Codes: 708032, 708036 | The two bolts that attach the clamping plate to the lateral shaft may break, which could cause th... | Class II | Philips Electronics North America Corporation |
| Jul 14, 2016 | Straumann Bone Level Implant, ¿ 4.1mm RC, SLA 10mm, TiZr, NTP Article Number... | Misalignment with the Loxim transfer piece markings, the positioning of the implant will be incor... | Class II | Straumann Manufacturing, Inc. |
| Jul 14, 2016 | Straumann Bone Level Implant, ¿ 4.1mm RC, SLA¿ 12mm, TiZr, NTP Article Numbe... | Misalignment with the Loxim transfer piece markings, the positioning of the implant will be incor... | Class II | Straumann Manufacturing, Inc. |
| Jul 7, 2016 | T2Candida¿Cartridges; Part Number: 90-01794 (box of 12) and Part Number: 91-0... | Elevated levels of invalid results. | Class II | T2 Biosystems Inc |
| Jul 7, 2016 | Allura Xper with release R8.2: Allura Xper FD10, 722026; Allura Xper FD10/10... | Increase in the failure rate of certain low-voltage DC power supplies (DCPS) | Class II | Philips Electronics North America Corporation |
| Jul 7, 2016 | UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD... | Increase in the failure rate of certain low-voltage DC power supplies (DCPS) | Class II | Philips Electronics North America Corporation |
| Jul 5, 2016 | Siemens RAPIDPoint¿ 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software ... | Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the ... | Class III | Siemens Healthcare Diagnostics Inc |
| Jun 24, 2016 | Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-steriliz... | Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inac... | Class II | PerkinElmer, Inc |
| Jun 22, 2016 | AnastoCiip Universal Clip Remover (1Each} REF# 4001-04 The Anastoclip Cl... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | AnastoCiip Universal Clip Remover (1Each} REF# e4001-04 The Anastoclip Cl... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | Anasto AC Closure System and Anasto Universal Clip Remover- Anasto AC 8cm 2... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument ... | Software defect, where either an Above Assay Range or an inaccurate value could potentially be re... | Class II | Siemens Healthcare Diagnostics Inc |
| Jun 22, 2016 | AnastoCiip Universal Clip Remover (6 Pack} REF# e4001-00 The Anastoclip ... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | Anastoclip AC 8cm 35 Clips (1 Each} Medium REF # 4000-06 The Anastoclip C... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | AnastoCiip Universal Clip Remover (6 Pack} REF# 4001-00 The Anastoclip ... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 22, 2016 | Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical ... | Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in tes... | Class II | Olympus Scientific Solutions Americas |
| Jun 22, 2016 | AnastoClip AC 8cm 35 Clips (1 Each} Large REF # 4000-07 The Anastoclip C... | US marketing brochure for the AnastoClip AC Closure System states that the Anastoclip catalog nu... | Class II | LeMaitre Vascular, Inc. |
| Jun 16, 2016 | AB SCIEX Triple Quad 4500MD LC/MS/MS System. Mass Spectrometer for In-Vitro... | MultiQuant MD software where under certain conditions a user can be presented with incorrect qua... | Class II | Ab Sciex |
| Jun 16, 2016 | AB SCIEX 3200MD QTRAP¿ LC/MS/MS System Mass Spectrometer for In-Vitro Diagn... | MultiQuant MD software where under certain conditions a user can be presented with incorrect qua... | Class II | Ab Sciex |
| Jun 16, 2016 | AB SCIEX QTRAP¿ 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagn... | MultiQuant MD software where under certain conditions a user can be presented with incorrect qua... | Class II | Ab Sciex |
| Jun 16, 2016 | AB Sciex API 3200MD" LC/MS/MS System. In-Vitro Diagnostic Instrument Part Nu... | MultiQuant MD software where under certain conditions a user can be presented with incorrect qua... | Class II | Ab Sciex |
| Jun 15, 2016 | Smith and Nephew TWINFIX QUICK-T Knot Pusher Suture Cutter; Part Number: 7209... | The device could be subject to breaches of its sterile packaging. | Class II | Smith & Nephew, Inc. |
| May 31, 2016 | ATEC CANISTER. Suction Canister with Lid 490 cc., RX Only; Suction canister u... | Cracks ATEC Canister Lids. | Class II | Hologic, Inc |
| May 24, 2016 | Crit Line in a Clip (CLiC). CliC P/N: CL 10041001 CLiC is a continuous real-... | Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine... | Class II | Fresenius Medical Care Renal Therapies Group, LLC |
| May 2, 2016 | Siemens Healthcare ADVIA Centaur Tnl-Ultra Assay . In-Vitro Diagnostic for qu... | Exhibits a greater than 10% change in results in samples with biotin levels up to 10 ng/mL (41 nm... | Class II | Siemens Healthcare Diagnostics, Inc |
| Apr 14, 2016 | HeartStart MRx Monitor/Defibrillator The HeartStart MRx is for use for the t... | MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may hav... | Class II | Philips Electronics North America Corporation |
| Mar 31, 2016 | Mass Eye & Ear Finger Cot Nasal 2's Ref Number: 41-04 Product Usage: Neu... | Product packaging defective compromising sterility | Class II | American Surgical Company |
| Mar 31, 2016 | American Surgical Tonsil Sponges 1" Large Ref Number: 50-01 Product Usage: ... | Product packaging defective compromising sterility | Class II | American Surgical Company |