DateProductReasonClassFirm
Dec 12, 2024 Impella RP Flex with SmartAssist; Product Number: 1000323; Optical Sensors have been damaged due to physical interaction between the inlet and another devic... Class I Abiomed, Inc.
Dec 12, 2024 SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburs... Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanicall... Class II LivaNova USA, Inc.
Dec 12, 2024 SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Ther... Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanicall... Class II LivaNova USA, Inc.
Dec 11, 2024 Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003... Spore test strip monitoring system, designed for routine monitoring of sterilization processes, m... Class II HF Acquisition Co LLC
Dec 11, 2024 MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.... Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent... Class II MICROVENTION INC.
Dec 11, 2024 nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extraco... The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and Fle... Class I Nuwellis Inc
Dec 10, 2024 Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic aci... False-positive MSI-H results generated by the Idylla MSI Test. Class II Biocartis Nv
Dec 10, 2024 Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Pa... Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized ... Class I Mercury Enterprises, Inc. dba Mercury Medical
Dec 10, 2024 Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W0... There is the potential for in-vivo balloon deflation issues, including difficulty or inability to... Class II ev3 Inc
Dec 10, 2024 EMPOWR 3D Knee Tibial Insert, 6R 10MM, VE, REF: 342-10-706 Knee insert package may contain an acetabular system, which may cause surgical delay to get a new... Class II Encore Medical, LP
Dec 9, 2024 Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model ... Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching fr... Class I Maquet Cardiovascular, LLC
Dec 9, 2024 VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-400... Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching fr... Class I Maquet Cardiovascular, LLC
Dec 9, 2024 JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L Oxygen concentrator may melt or catch on fire during use. Class I JIANGSU JUMAO X-CARE MEDICAL EQUIPMENT CO LTD
Dec 6, 2024 Temporary Titanium Abutments Certain definitive dental implant restoration cases may have included a temporary titanium abutme... Class II DDS Lab
Dec 6, 2024 Portrait Core Services, a set of software services that enable the communicat... There is a software issue that can result in the loss of patient monitoring on the Portrait Centr... Class II GE Medical Systems, LLC
Dec 6, 2024 Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004 Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could ... Class I Fresenius Kabi USA, LLC
Dec 5, 2024 Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Product... Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cau... Class II QUIDEL ORTHO
Dec 5, 2024 Dialyzer Optiflux 160NRe Potential for internal blood leaks due to cracked polyurethane Class II Fresenius Medical Care Holdings, Inc.
Dec 5, 2024 LINK MP Reconstruction System Neck Segment (with suture holes, standard, unc... Error on the main label and the patient label. The labels indicate a size XXL, this is a standard... Class II Waldemar Link GmbH & Co. KG (Mfg Site)
Dec 5, 2024 Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Product... Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cau... Class II QUIDEL ORTHO
Dec 4, 2024 Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Contr... Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and T... Class II Augustine Temperature Management, LLC
Dec 4, 2024 Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of... Incorrect catheter connector within the kit resulting in inability to provide a secure connection... Class II B. Braun Medical, Inc.
Dec 4, 2024 Hillrom Temperature Management Controller, MP, Model # 2083516; used in conju... Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and T... Class II Augustine Temperature Management, LLC
Dec 4, 2024 Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Contr... Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and T... Class II Augustine Temperature Management, LLC
Dec 4, 2024 Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150M... Due to an increase in complaint that the percutaneous reference pin would not fit into the patien... Class II Medtronic Navigation, Inc.
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE IUD EXTRACT... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE CEMENT SPAT... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-... Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacti... Class II Stryker Corporation
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE MINI TIS... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE IRIS SCI... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE BLAKE GALLS... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE IRIS HOOK Q... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE SCHAMBERG E... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE brand medical procedure kits labeled as: 1) ECONO STERILE" B... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE SIMS UTERIN... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VEIN HOO... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE D/A NAIL... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE SCALPEL HDL... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 Medical procedure kits labeled as: 1) ECONO STERILE STRABIS SCISS STR 4.5"... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE GRAVES... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RET... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE LEMPERT BON... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE brand medical procedure kits labeled as: 1) ECONO STERILE FR... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE PROBE W/... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE WEITLANE... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE KOGAN ENDOC... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE OLSEN-HE... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE YANKAUER... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: ECONO STERILE ALFONSO INF... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments
Dec 3, 2024 ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VIENNA N... Reports of various packaging issues that may result in a breach of the sterile barrier. Class II Sklar Instruments

Frequently Asked Questions

When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.

All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.