Portrait Core Services, a set of software services that enable the communication and interaction ...
FDA Device Recall #Z-0918-2025 — Class II — December 6, 2024
Recall Summary
| Recall Number | Z-0918-2025 |
| Classification | Class II — Moderate risk |
| Date Initiated | December 6, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Medical Systems, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 7 systems |
Product Description
Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB
Reason for Recall
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Lot / Code Information
UDI-DI: 00195278659446, Serial Numbers: SWT24040001TA SWT24090006TA SWT24150008TA SWT24050001TA SWT24180011TA SWT24180009TA SWT31240012TA SWT24340007TA SWT24360002TA SWT24370001TA SWT24370002TA SWT24180010TA 6399074-7 6399074-8 6399074-9 6399326-1 6399326-2 SWT24060001TA
Other Recalls from GE Medical Systems, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2154-2026 | Class II | Revolution Apex, X-ray Computed Tomography, Mod... | Mar 26, 2026 |
| Z-2153-2026 | Class II | Revolution CT ES, X-ray Computed Tomography, Mo... | Mar 26, 2026 |
| Z-2149-2026 | Class II | Revolution Ascend, computed tomography, Model N... | Mar 26, 2026 |
| Z-2151-2026 | Class II | Revolution Apex Elite, X-ray Computed Tomograph... | Mar 26, 2026 |
| Z-2150-2026 | Class II | Revolution CT, x-ray computed tomography, Model... | Mar 26, 2026 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?