Portrait Core Services, a set of software services that enable the communication and interaction ...
FDA Recall #Z-0918-2025 — Class II — December 6, 2024
Product Description
Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB
Reason for Recall
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
Recalling Firm
GE Medical Systems, LLC — Waukesha, WI
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
7 systems
Distribution
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Code Information
UDI-DI: 00195278659446, Serial Numbers: SWT24040001TA SWT24090006TA SWT24150008TA SWT24050001TA SWT24180011TA SWT24180009TA SWT31240012TA SWT24340007TA SWT24360002TA SWT24370001TA SWT24370002TA SWT24180010TA 6399074-7 6399074-8 6399074-9 6399326-1 6399326-2 SWT24060001TA
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated