MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 1...
FDA Recall #Z-1121-2025 — Class II — December 11, 2024
Product Description
MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
Reason for Recall
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
Recalling Firm
MICROVENTION INC. — Aliso Viejo, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
11 units
Distribution
US: None OUS: China
Code Information
Lot number: 0000456768
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.