| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 350010-1, 350000-2, 350010-X | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB ... | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (H... | Class II | Beckman Coulter Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 560010-2L, 560010-XL, 560010-1L, 56... | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 860010-1L | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 361910-1A, 361910-1A, 361910-XA, 36... | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 360010-2WT, 360010-1, 370000-1W, 36... | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB... | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (H... | Class II | Beckman Coulter Inc. |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, ... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 760010-1, 770010-1 | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Brilliance 16 Power, Model... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB ... | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (H... | Class II | Beckman Coulter Inc. |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Big Bore RT, Model Number:... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Nu... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Philips CT systems labeled as the following: 1. Brilliance iCT, Model Numb... | The patient support table (couch) may descend unexpectedly to the lowermost position due to a com... | Class II | Philips North America Llc |
| Aug 13, 2025 | Change Healthcare Radiology Solutions software version 14.2.2 | Due to software issue, radiology reports may not be fully displayed when viewing. | Class II | CHANGE HEALTHCARE CANADA COMPANY |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 360030-1C, 360030-1CE, 360030-1CTS,... | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 360020-1EP | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 13, 2025 | UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB ... | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (H... | Class II | Beckman Coulter Inc. |
| Aug 13, 2025 | Penner Pacific Bathing Spa, Model Numbers 360020-1P | The device does not bear a unique device identifier. | Class III | Penner Patient Care, Inc. |
| Aug 12, 2025 | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 90, REF E NR... | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at l... | Class II | Applied Medical Technology Inc |
| Aug 12, 2025 | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 90, REF E NR... | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at l... | Class II | Applied Medical Technology Inc |
| Aug 12, 2025 | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NR... | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at l... | Class II | Applied Medical Technology Inc |
| Aug 12, 2025 | Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use in... | Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, th... | Class I | Max Mobility LLC |
| Aug 12, 2025 | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 55, REF E NR... | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at l... | Class II | Applied Medical Technology Inc |
| Aug 12, 2025 | Artis Pheno. Image-Intensified Flouroscopic X-Ray System. | Limited system movements after startup . | Class II | Siemens Medical Solutions USA, Inc |
| Aug 12, 2025 | Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in ... | Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, th... | Class I | Max Mobility LLC |
| Aug 12, 2025 | Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use i... | Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, th... | Class I | Max Mobility LLC |
| Aug 11, 2025 | Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 1109... | Potential for falsely depressed patient, quality control (QC), and/or calibration results or dela... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 11, 2025 | Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions ... | Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schill... | Class II | Merge Healthcare, Inc. |
| Aug 11, 2025 | Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use ... | Their is the potential for weak staining which may result in false negative CD20 identification. | Class II | Agilent Technologies Denmark ApS |
| Aug 11, 2025 | Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use ... | Their is the potential for weak staining which may result in false negative CD20 identification. | Class II | Agilent Technologies Denmark ApS |
| Aug 11, 2025 | Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use ... | Their is the potential for weak staining which may result in false negative CD20 identification. | Class II | Agilent Technologies Denmark ApS |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 2023B, ... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 7, Soft... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322 | Blood collection tubes were manufactured with the label applied too high on the tube, which shif... | Class II | Greiner Bio-One North America, Inc. |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 12A, So... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 10B, So... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | Sterile Medline Convenience Kits: 1) MAJOR PACK L-F, Model Number: DYNJ03827... | Convenience kits labeled as sterile have not gone through the sterilization process. | Class I | Medline Industries, LP |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 6 Servi... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 9B, Sof... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 11A, So... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | Sterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B | Convenience kits labeled as sterile have not gone through the sterilization process. | Class I | Medline Industries, LP |
| Aug 8, 2025 | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.9... | Vital sign monitoring instrument may trigger an error message during blood pressures measurement... | Class II | Schiller, Ag |
| Aug 8, 2025 | IntelliSpace Cardiovascular, Software 8.0.0.4. | Software issue that results in the display of outdated information. | Class II | Philips Medical Systems Nederland B.V. |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 8B, Sof... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | RayStation with the following product descriptions: 1. RayStation 11B, So... | DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's i... | Class II | RAYSEARCH LABORATORIES AB |
| Aug 8, 2025 | Baster Continu-Flo Solution Set, Non-Vented, 2 Clearlink Luer Activated Valve... | Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-F... | Class II | Baxter Healthcare Corporation |