| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jun 29, 2018 | Pure Rayz, Light Therapy System, packaging, box, sleeve and user manual. Hand... | Quasar did not receive FDA clearance for claims of collagen/elastin production as stated on the p... | Class II | Quasar Bio-Tech, Inc. |
| Jun 29, 2018 | Atellica Sample Handler Additional (SMN 11069004). The Atellica" Solution ... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Dymaxeon 5.5 Curved Rod Size 5.5 x 110mm (Yellow), Model Number 06.6441B P... | Packages labeled as containing 110mm rods may contain 100mm rods. | Class II | Back 2 Basics Direct LLC |
| Jun 29, 2018 | Atellica IM 1600 Analyzer (SMN 11066000). The Atellica" Solution is a multi-... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | MAYFIELD SPINE TABLE ADAPTOR RADIOLUCENT, Part Number A2600R The MAYFIELD... | There is a potential for the knob assembly in the swivel adapter to fracture/break during use. | Class II | Integra LifeSciences Corp. |
| Jun 29, 2018 | Atellica Sample Handler Prime (SMN 11069001). The Atellica" Solution is a ... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 28, 2018 | Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Cat... | The torque wrench used on the ceiling column assembly during a recent service visit was outside o... | Class II | Bayer Medical Care, Inc. |
| Jun 28, 2018 | Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Cat... | The torque wrench used on the ceiling column assembly during a recent service visit was outside o... | Class II | Bayer Medical Care, Inc. |
| Jun 28, 2018 | The Neonatal Flow Sensor Cable used on devices CARESCAPE R860, ENGSTROM CARES... | The Neonatal Flow Sensor Cable may be missing the connector housing exposing the wires in the cable. | Class II | Datex-Ohmeda, Inc. |
| Jun 28, 2018 | Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with... | The torque wrench used on the ceiling column assembly during a recent service visit was outside o... | Class II | Bayer Medical Care, Inc. |
| Jun 28, 2018 | Final Pack, Esteem II, Sound Processor Model 2001, containing 2 Sterile Wipes... | Sterile Wipes packed in these four Esteem II Sound Processors had an earlier expiration date than... | Class II | Envoy Medical Corporation |
| Jun 27, 2018 | DigitalDiagnost Rel.4 Upgrades (Stitching Patient Support) 712084 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4.1 High Performance (Stitching Patient Support) 712031 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | BACT/ALERT VIRTUO system, A Unit, with Version R2.0 Firmware Product Usage... | The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident... | Class II | bioMerieux, Inc. |
| Jun 27, 2018 | DigitalDiagnost, Dual-Detector (Stitching Patient Support) 712052 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4 Flex / Value (Stitching Patient Support) 712028 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | Proteus 235, Universal Nozzle with snout 300x400; PBS dedicated nozzle with s... | Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not cor... | Class II | Ion Beam Applications S.A. |
| Jun 27, 2018 | DigitalDiagnost Dual Detector (Stitching Patient Support) 712022 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | Stryker T2 ANKLE ARTHRODESIS NAIL RIGHT 10MMX200MM Item Number: 1819-1020S T... | Incomplete seal on the Tyvek lid may compromise the sterility of the device | Class II | Stryker GmbH |
| Jun 27, 2018 | DigitalDiagnost Classic, Dual-Detector (Stitching Patient Support) 712052 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | CombiDiagnost R90 (Stitching Patient Support) 709030 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Classic 1.x Upgrades (Stitching Patient Support) 712081 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4.1 (Stitching Patient Support) 712225 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | EasyDiagnost Eleva DRF, Release 5 (Stitching Patient Support) 706050 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4 Chest / Emergency (Stitching Patient Support) 712029 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 2.x Upgrades (Stitching Patient Support) 712082 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost, Single-Detect (Stitching Patient Support) 712062 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | BACT/ALERT VIRTUO system, A Unit China, with Version R2.0 Firmware Product... | The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident... | Class II | bioMerieux, Inc. |
| Jun 27, 2018 | DigitalDiagnost 4.1 Chest / Emergency (Stitching Patient Support) 712029 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Upgrades (Stitching Patient Support) 712083 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device i... | Genicon is providing updated IFU labeling to users to provide clarity on how to use the device fo... | Class II | Genicon, Inc. |
| Jun 27, 2018 | DigitalDiagnost Release 3 (Stitching Patient Support) 712025 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | Leksell GammaPlan 11.1 Product Usage: Leksell GammaPlan is a computer-bas... | The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of... | Class II | Elekta Instrument AB |
| Jun 27, 2018 | BACT/ALERT VIRTUO system, B Unit, with Version R2.0 Firmware Product Usage... | The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident... | Class II | bioMerieux, Inc. |
| Jun 27, 2018 | DigitalDiagnost 2.x Upgrades (Stitching Patient Support) 712081 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Release 3 (Stitching Patient Support) 712025 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4 High Performance (Stitching Patient Support) 712027 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4.1 Flex / Value (Stitching Patient Support) 712028 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4 Flex / Value (Stitching Patient Support) 712028 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | EasyDiagnost Eleva DRF (Stitching Patient Support) 706032 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Single Detector (Stitching Patient Support) 712020 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost 4.1 High Performance (Stitching Patient Support) 712031 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | DigitalDiagnost Dual Detector (Stitching Patient Support) 712022 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 27, 2018 | BuckyDiagnost Ceiling System (Stitching Patient Support) 704031 | The hook does not securely hold the footplate in the vertical position. Therefore, the footplate... | Class II | Philips Medical Systems Gmbh, DMC |
| Jun 26, 2018 | Instrument trays used to store AltiVate Small Shell Trials (FA S ALTIVATE RSP... | A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide ... | Class II | Encore Medical, Lp |
| Jun 26, 2018 | Monitor with one Blood Parameter Module and one HSat Probe; 500AHCT Monitor ... | H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulte... | Class II | Terumo Cardiovascular Systems Corporation |
| Jun 26, 2018 | CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product U... | H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulte... | Class II | Terumo Cardiovascular Systems Corporation |
| Jun 25, 2018 | This is an accessory to Intelect Advanced, Intelect Mobile, Intelect Neo, and... | The device accessories were not labeled with appropriate latex warning. | Class II | DJO, LLC |