Browse Device Recalls
3,013 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 3,013 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 3,013 FDA device recalls in 2018.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 125¿ 13 mm x 260 mm, Item Number: 814413260 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 125¿ 9 mm x 360 mm, Item Number: 814409360 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail 125¿ 9 mm x 180 mm, Item Number: 814309180 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 15 mm x 360 mm, Item Number: 814515360 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 130¿ 13 mm x 280 mm, Item Number: 814613280 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 9 mm x 380 mm, Item Number: 814509380 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 125¿ 9 mm x 300 mm, Item Number: 814409300 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 9 mm x 360 mm, Item Number: 814509360 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 130¿ 11 mm x 400 mm, Item Number: 814611400 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 15 mm x 400 mm, Item Number: 814515400 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 130¿ 11 mm x 360 mm, Item Number: 814611360 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 11 mm x 280 mm, Item Number: 814511280 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Left 125¿ 11 mm x 360 mm, Item Number: 814411360 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 130¿ 13 mm x 460 mm, Item Number: 814513460 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 125¿ 9 mm x 460 mm, Item Number: 814309460 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 17, 2018 | Affixus¿ Hip Fracture Nail Right 125¿ 9 mm x 440 mm, Item Number: 814309440 | The set screw is not able to advance or reverse from the original position in the nail. | Class II | Zimmer Biomet, Inc. |
| Jul 16, 2018 | Allura Xper F010, System Code 722003 Vascular, cardiovascular and neurova... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper FD2O Biplane, System Code 722008 Vascular, cardiovascular and... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | AIIuraXperFDlO, System Code 722010 Vascular, cardiovascular and neurovasc... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | INTEGRIS Allura 9 F FDXD, System Code 722497 Vascular, cardiovascular and... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | INTEGRIS Allura 15-12 (biplane), System Code 722044 Vascular, cardiovascu... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | RAPIDPoint 405 Blood Gas Analyzer; 10282093, 10310464, 10314817, 10317193, 10... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | AlluraXperFOlO/lO, System Code 722005 Vascular, cardiovascular and neurov... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Cesar-DMCP-Visub(HM2000/3000), System Code 72239 Vascular, cardiovascula... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | CoLink(TM) Lapidus Plate XP, +2 mm, Left, REF P40 ST265, QTY 1, In2Bones ... | CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, righ... | Class II | In2bones USA, LLC |
| Jul 16, 2018 | RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306 ... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | CoLink(TM) Lapidus Plate XP, +2 mm, Right, REF P40 ST165, QTY 1, In2Bones ... | CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, righ... | Class II | In2bones USA, LLC |
| Jul 16, 2018 | Allura Xper F020 Biplane DR Table, System Code 722020 Vascular, cardiovas... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper F020 DR Table, System Code 722023 Vascular, cardiovascular an... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Valeris Medical Apollo Medial Suture Anchor, Catalog Number MSA-4515, 4.5 x 1... | The product was mislabeled as containing the incorrect type of suture. | Class II | Valeris Medical, LLC |
| Jul 16, 2018 | Cesar Powerpack-Visub(V3000), System Code 72243 Vascular, cardiovascular ... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | da Vinci Si/X/Xi Surgical System Redundant Medical Grade Power Supply (RMGPS)... | Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patien... | Class II | Intuitive Surgical, Inc. |
| Jul 16, 2018 | Allura Xper F010 DR Table, System Code 722014 Vascular, cardiovascular an... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper F020 DR Table, System Code 722015 Vascular, cardiovascular an... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | INTEGRIS Allura 9, System Code 722018 Vascular, cardiovascular and neurov... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | AlIura Xper F010 F, System Code 722002 Vascular, cardiovascular and neur... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper FD2O Biplane, System Code 722013 Vascular, cardiovascular and... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cob... | Sample material may potentially come in contact with the pipetting nozzle during operation of cob... | Class II | Roche Diagnostics Corporation |
| Jul 16, 2018 | Allura Xper F010/10 DRTable, System Code 722019 Vascular, cardiovascular ... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | NTEGRIS Allura 9 0 FDXD, System Code 722498 Vascular, cardiovascular and... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | INTEGRIS Allura 15-12 (mono), System Code 722043 Vascular, cardiovascular... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper FD2O, System Code 722012 Vascular, cardiovascular and neurova... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is ... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | INTEGRIS Allura 9 (biplane), System Code 722021 Vascular, cardiovascular ... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurova... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | AIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular an... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | INTEGRIS SUITE, System Code 722199 Vascular, cardiovascular and neurovasc... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Poly C- DMCP-Visub(H3000), System Code 72238 Vascular, cardiovascular and... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
| Jul 16, 2018 | Allura Xper FD1O C, System Code 722001 Vascular, cardiovascular and neuro... | There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may... | Class II | Philips Electronics North America Corporation |
Frequently Asked Questions
When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.
All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.
Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.