Medical Device Recalls
Browse 3,099 FDA device recalls.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jul 14, 2022 | cobas e801 Immunoassay Analyzer | Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The... | Class II | Roche Diagnostics Operations, Inc. |
| Jul 14, 2022 | Aeos¿ Robotic Digital Microscope with software versions 2.7 or earlier Artic... | Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recov... | Class II | Aesculap Implant Systems LLC |
| Jul 13, 2022 | VITEK 2 / MYLA VITEK 2 Software versions 8.01, 8.02 (veterinary use only),... | There are 7 reported software anomalies that may affect use of the device. | Class II | bioMerieux, Inc. |
| Jul 12, 2022 | Surpria: Software Version V3.11, V3.22 | Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstruc... | Class II | FUJIFILM Healthcare Americas Corporation |
| Jul 12, 2022 | SCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C | Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstruc... | Class II | FUJIFILM Healthcare Americas Corporation |
| Jul 12, 2022 | Surpria 64: Software Version V3.11, V3.22 | Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstruc... | Class II | FUJIFILM Healthcare Americas Corporation |
| Jul 8, 2022 | Pentacam HR REF 70900. Used to image the anterior segment of the eye. | Due to a software issue, IOL calculator printout often does not accurately reflect the alignment ... | Class II | Oculus Optikgeraete GMBH |
| Jul 8, 2022 | Pentacam AXL Wave REF 70020. Used to image the anterior segment of the eye. | Due to a software issue, IOL calculator printout often does not accurately reflect the alignment ... | Class II | Oculus Optikgeraete GMBH |
| Jul 8, 2022 | Pentacam AXL REF 70100, Pentacam HR REF 70900. Used to image the anterior... | Due to a software issue, IOL calculator printout often does not accurately reflect the alignment ... | Class II | Oculus Optikgeraete GMBH |
| Jul 7, 2022 | The Programmer/Recorder/Monitor (PRM) is intended to be used as part of the Z... | There is potential for unanticipated software anomalies to occur when using the Model 2868 Applic... | Class II | Boston Scientific Corporation |
| Jul 7, 2022 | The PRM is intended to be used as part of the ZOOM LATITUDE Programming Syste... | There is potential for unanticipated software anomalies to occur when using the Model 2868 Applic... | Class II | Boston Scientific Corporation |
| Jul 7, 2022 | The PRM is intended to be used as part of the ZOOM LATITUDE Programming Syste... | There is potential for unanticipated software anomalies to occur when using the Model 2868 Applic... | Class II | Boston Scientific Corporation |
| Jul 5, 2022 | ZAP-X Radiosurgery System | Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to ... | Class II | Zap Surgical Systems |
| Jun 23, 2022 | INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) ... | The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible ... | Class II | Mallinckrodt, LLC. |
| Jun 23, 2022 | Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and an... | A race condition between the storage system and services database has the potential to revert edi... | Class II | DICOM Grid, Inc. d/b/a Ambra Health |
| Jun 22, 2022 | ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: ... | There is a potential risk of medication error when using specific Abacus Order Entry & Calculatio... | Class I | Baxter Healthcare Corporation |
| Jun 17, 2022 | Merlin.net model MN5000 Software | The Merlin" 2 PCS model MER3700 is a portable, dedicated programming system designed to interrog... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Jun 17, 2022 | Merlin PCS 3650 programmer Model 3330 software | There is a potential for programmer software (Merlin PCS) and remote monitoring software applicat... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Jun 17, 2022 | Merlin 2 PCS MER37000 programmer Model MER3400 software | The Merlin" 2 PCS model MER3700 is a portable, dedicated programming system designed to interrog... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Jun 16, 2022 | nordicBrainEX, software versions 2.21 through 2.3.10 | When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both th... | Class II | NordicNeuroLab AS |
| Jun 15, 2022 | Sensis Vibe Combo with software version VD12A. A diagnostic and administrati... | Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication du... | Class II | Siemens Medical Solutions USA, Inc |
| Jun 15, 2022 | Sensis with software version VD12A. A diagnostic and administrative tool su... | Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication du... | Class II | Siemens Medical Solutions USA, Inc |
| Jun 15, 2022 | SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-... | When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge t... | Class II | Stryker Instruments Div. of Stryker Corporation |
| Jun 14, 2022 | Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1 | The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and ... | Class II | SIGHT DIAGNOSTICS LTD |
| Jun 14, 2022 | Centricity PACS Software Version 7.0 SP0.0.4.7 | The Event Notification Manager (ENM) functionality for certain products does not process notifica... | Class II | GE Healthcare, LLC |
| Jun 2, 2022 | Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilat... | A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into... | Class II | Breas Medical, Inc. |
| May 31, 2022 | NaviCare Nurse Call/Voalte Nurse Call | An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, C... | Class II | Baxter Healthcare Corporation |
| May 31, 2022 | Xhibit Telemetry Receiver, Model: 96280 | Software related alarm escalation defect occurs after approximately 25 days of continuous use. Te... | Class II | Spacelabs Healthcare, Inc. |
| May 24, 2022 | t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insu... | Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate ... | Class II | Tandem Diabetes Care Inc |
| May 24, 2022 | BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with ... | The automated dispensing cabinet software update included a .dll file that is not compatible with... | Class II | CareFusion 303, Inc. |
| May 24, 2022 | t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insu... | Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate ... | Class II | Tandem Diabetes Care Inc |
| May 3, 2022 | NIM Vital Console 4.0 (P/N: NIM4CM01) NIM with Software version 1.3.2 | Software anomaly was identified. | Class II | Medtronic Xomed, Inc. |
| May 3, 2022 | NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2 | Software anomaly was identified. | Class II | Medtronic Xomed, Inc. |
| Apr 26, 2022 | 4 Series CX phototherapy units equipped with Daavlin's ClearLink Control Syst... | Software issue, resuming an interrupted treatment will result in swap of utilized calibration tre... | Class II | Daavlin Distributing Company |
| Apr 26, 2022 | 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control Syst... | Software issue, resuming an interrupted treatment will result in swap of utilized calibration tre... | Class II | Daavlin Distributing Company |
| Apr 25, 2022 | CardioTek EP-TRACER Software V2.x. | Device did not pass electrical safety testing for adequate insulation. | Class II | CardioTek BV |
| Apr 22, 2022 | Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0... | Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed t... | Class II | Fresenius Kabi USA, LLC |
| Apr 21, 2022 | ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1... | Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may... | Class II | Dain Technology, Inc. (Subsidiary of Soundable ... |
| Apr 19, 2022 | Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software ... | When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or... | Class II | Haemonetics Corporation |
| Apr 18, 2022 | Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-2... | Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audi... | Class I | Smiths Medical ASD Inc. |
| Apr 13, 2022 | Sensis, Programmable Diagnostic Computer, Model Nos. 6623974 ... | Siemens has become aware of three potential software issues with AXIOM Sensis or Sensis/ Sensis L... | Class II | Siemens Medical Solutions USA, Inc |
| Apr 12, 2022 | Cobalt Models loaded with CareLink SmartSync Device Manager application softw... | Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrill... | Class II | Medtronic Inc., Cardiac Rhythm and Heart Failur... |
| Apr 12, 2022 | Crome Models loaded with CareLink SmartSync Device Manager application softwa... | Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrill... | Class II | Medtronic Inc., Cardiac Rhythm and Heart Failur... |
| Apr 7, 2022 | Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reor... | The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Mod... | Class II | Bio-Rad Laboratories, Inc. |
| Mar 29, 2022 | The device is a whole-body CT x-ray system. The acquired x-ray transmission d... | Three software issues affecting incorrect image display, error interpreting patient images due to... | Class II | Philips North America Llc |
| Mar 29, 2022 | VITEK 2 Systems and VITEK 2 with MYLA. | Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and syner... | Class II | bioMerieux, Inc. |
| Mar 21, 2022 | ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0... | Software issue for treatment plan and image management application may result in mismatch values ... | Class II | Varian Medical Systems Imaging Laboratory GmbH |
| Mar 15, 2022 | PrisMax System, Product Code 955626 | This correction is due to software anomalies occurring during use. If the operator initiates ther... | Class II | Baxter Healthcare Corporation |
| Mar 15, 2022 | PrisMax System, Product Code 955724 | This correction is due to software anomalies occurring during use. If the operator initiates ther... | Class II | Baxter Healthcare Corporation |
| Mar 15, 2022 | PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701 | This correction is due to software anomalies occurring during use. If the operator initiates ther... | Class II | Baxter Healthcare Corporation |