| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Dec 21, 2018 | ID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/... | Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit ... | Class I | Draeger Medical, Inc. |
| Dec 21, 2018 | VentStar Basic 250, Catalog Number MP00350 Breathing Circuits/Anesthesia S... | Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit ... | Class I | Draeger Medical, Inc. |
| Dec 21, 2018 | VentStar Watertrap (P) 180, Catalog Number MP00361 Breathing Circuits/Ane... | Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit ... | Class I | Draeger Medical, Inc. |
| Dec 21, 2018 | ID Circuit Basic 180, Catalog Number MP01350 Breathing Circuits/Anesthesi... | Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit ... | Class I | Draeger Medical, Inc. |
| May 24, 2018 | Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a... | The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC mo... | Class II | Draeger Medical, Inc. |
| May 24, 2018 | Fabius GS MRI Anesthesia Machine; Cat. no. 8607300 Product Usage: Inhalat... | The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC mo... | Class II | Draeger Medical, Inc. |
| May 24, 2018 | Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inh... | The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC mo... | Class II | Draeger Medical, Inc. |
| May 24, 2018 | Fabius GS Tiro Anesthesia Machine; Cat. no. 8606000 Product Usage: Inhala... | The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC mo... | Class II | Draeger Medical, Inc. |
| May 24, 2018 | Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usa... | The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC mo... | Class II | Draeger Medical, Inc. |
| Apr 26, 2018 | Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator f... | The valve assembly material may become corroded due to insufficient cleaning by a supplier. There... | Class II | Draeger Medical, Inc. |
| Mar 1, 2018 | Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The... | The automatic ventilation may fail if the position detection of the motor is disturbed. If the ve... | Class I | Draeger Medical, Inc. |
| Mar 1, 2018 | Fabius Tiro, anesthesia machine, catalog no. 8606000 Product Usage: The p... | The automatic ventilation may fail if the position detection of the motor is disturbed. If the ve... | Class I | Draeger Medical, Inc. |
| Mar 1, 2018 | Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: ... | The automatic ventilation may fail if the position detection of the motor is disturbed. If the ve... | Class I | Draeger Medical, Inc. |
| Mar 1, 2018 | Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The pr... | The automatic ventilation may fail if the position detection of the motor is disturbed. If the ve... | Class I | Draeger Medical, Inc. |
| Jun 22, 2016 | Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog #... | Risk associated with operating the device in skin temperature mode with temperature sensors attac... | Class II | Draeger Medical, Inc. |
| May 31, 2016 | Oxylog 3000, Disposable Pediatric Patient Circuit, Catalog Number: 5704964 ... | Isolated episodes of leakage at the check valve were found | Class I | Draeger Medical, Inc. |
| Apr 21, 2016 | Draeger D-Vapor/ D-Vapor 3000 Vaporizers with Baxter Desflurane Agent and Bot... | Operators noticed unwanted release of evaporating anesthetic agent at the end of the filling proc... | Class II | Draeger Medical, Inc. |
| Feb 4, 2016 | Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation. | The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation C... | Class I | Draeger Medical, Inc. |
| Dec 22, 2015 | Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus V... | Drager became aware of situations where the error message "Poti unplugged" was generated causing ... | Class I | Draeger Medical, Inc. |
| Dec 22, 2015 | Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: ... | Drager became aware of situations where the error message "Poti unplugged" was generated causing ... | Class I | Draeger Medical, Inc. |
| Dec 1, 2015 | Optional PS500 Power Supply Unit for the Evita V500 Ventilator and Babylog VN... | The firm became aware of cases in which the battery run times of the optional PS 500 power supply... | Class I | Draeger Medical, Inc. |
| Nov 12, 2015 | Perseus A500 Anesthesia Workstation, MK06000. Intended for use in anesthet... | There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Wor... | Class I | Draeger Medical, Inc. |
| Jul 13, 2015 | PS500 of an IACS Workstation Critical Care (Evita Infinity V500); intended fo... | The battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V... | Class I | Draeger Medical, Inc. |
| Jun 22, 2015 | Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine... | the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fab... | Class II | Draeger Medical, Inc. |
| Mar 25, 2014 | Reusable latex breathing bags with multiple part numbers and sold as part of ... | Reusable latex breathing bags with various part numbers and sold as part of test lung and reusabl... | Class II | Draeger Medical, Inc. |
| Sep 30, 2013 | Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in ... | Cases were encountered with the Draeger Movita 603x/DVE 803x Ceiling Supply Units where the drive... | Class II | Draeger Medical, Inc. |
| Aug 14, 2013 | Draeger Fabius GS Premium, Fabius OS, Fabius Tiro, Fabius Tiro D-M Product... | During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia mach... | Class I | Draeger Medical, Inc. |
| Dec 13, 2012 | Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number... | The displayed value for the invasive blood pressure measurement of the IACS was different than th... | Class II | Draeger Medical, Inc. |