Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity M...

FDA Recall #Z-0939-2013 — Class II — December 13, 2012

Recall #Z-0939-2013 Date: December 13, 2012 Classification: Class II Status: Terminated

Product Description

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Reason for Recall

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Recalling Firm

Draeger Medical, Inc. — Telford, PA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

260

Distribution

Distributed in the states of CA, IL, and MA.

Code Information

Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated