Browse Device Recalls
1,062 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 1,062 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 1,062 FDA device recalls in OH.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Mar 16, 2015 | MX8000 Dualv. EXP, 728130. Computed Tomography X-ray systems intended to p... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Ingenuity CT. Computed Tomography X-ray systems intended to produce cross-... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT Big Bore Oncology. Computed Tomography X-ray systems intende... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 64. Computed Tomography X-ray systems intended to produce cr... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 6 Air. Computed Tomography X-ray systems intended to produce... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 16 Air. Computed Tomography X-ray systems intended to produc... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 10 Air. Computed Tomography X-ray systems intended to produc... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 16 Power. Computed Tomography X-ray systems intended to prod... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance iCT. Computed Tomography X-ray systems intended to produce cros... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 16, 2015 | Brilliance CT 16 Water. Computed Tomography X-ray systems intended to prod... | The firm was informed that while raising the patient couch on the system to perform an exam, the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 13, 2015 | Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, AD... | During clinical use of a Vertex Plus Gamma Camera, as the detector heads were in the relative 180... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 11, 2015 | Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imagin... | It was noticed that the spectroscopy voxel map did not align with the anatomical images of the ph... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 11, 2015 | Xpress System Pedicle Screw Assembly, 5.5 x 55mm. An internal fixation devi... | One lot of instruments have etched on the Yoke of the Pedicle Screw a laser marking size of 7.5 x... | Class II | X Spine Systems Inc |
| Mar 11, 2015 | Nuclear gamma cameras (ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60 & ADAC ... | The firm was notified by a customer that the collimator exchange carriage and the collimator stor... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 10, 2015 | Talon Grasping Device, 160 cm, US endoscopy. Used for retrieval of foreign b... | A wire component on the distal grasping assembly of the device became detached.. | Class II | US Endoscopy Group Inc |
| Mar 6, 2015 | Brilliance CT 16 Slice Air (Uninterruptible Power Supply (UPS) accessory devi... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Precedence 6 (Uninterruptible Power Supply (UPS) accessory devices) Computed ... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity TF (Uninterruptible Power Supply (UPS) accessory devices) Computed ... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | BrightView X (Uninterruptible Power Supply (UPS) accessory devices) Computed ... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | CardioMD III (Uninterruptible Power Supply (UPS) accessory devices) Computed ... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Diamond Select Brilliance CT 16 (Uninterruptible Power Supply (UPS) accessory... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | GEMINI TF PET/CT 16 Slice (Uninterruptible Power Supply (UPS) accessory devic... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance CT 40 - Slices (Uninterruptible Power Supply (UPS) accessory devic... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Forte-JetStream AZ (Uninterruptible Power Supply (UPS) accessory devices) Com... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | iCT SP (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomogr... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance CT 64 channel with Essence, Technology (Uninterruptible Power Supp... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Com... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | GEMINI TF Ready (Uninterruptible Power Supply (UPS) accessory devices) Comput... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | BrightView XCT (Uninterruptible Power Supply (UPS) accessory devices) Compute... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance CT Big Bore Radiology (Uninterruptible Power Supply (UPS) accessor... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | AXIS (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomograp... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance CT 40 - Slices(Uninterruptible Power Supply (UPS) accessory device... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Gemini 16 PET/CT System (Uninterruptible Power Supply (UPS) accessory device... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Extended Brilliance Workstation (Uninterruptible Power Supply (UPS) accessory... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Precedence 16 (Uninterruptible Power Supply (UPS) accessory devices) Computed... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity CT (Uninterruptible Power Supply (UPS) accessory devices) Computed ... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | BrightView SPECT (Uninterruptible Power Supply (UPS) accessory devices) Compu... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | TruFlight Select PET/CT 16-slice (Uninterruptible Power Supply (UPS) accesso... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | FORTE (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomogra... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Gemini GXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) C... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance iCT (Uninterruptible Power Supply (UPS) accessory devices) Compute... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity Core (Uninterruptible Power Supply (UPS) accessory devices) Compute... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Compute... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Brilliance CT 6 Slice Air (Uninterruptible Power Supply (UPS) accessory devic... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Ingenuity PET/MR (Uninterruptible Power Supply (UPS) accessory devices) Comp... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | GEMINI TF Big Bore (Uninterruptible Power Supply (UPS) accessory devices) Com... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | Gemini LXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) C... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devic... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | MX 16 Slice Phase II (Uninterruptible Power Supply (UPS) accessory devices) C... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 6, 2015 | MX 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed T... | The firm was notified that the failure of the UPS accessory devices used in conjunction with Comp... | Class II | Philips Medical Systems (Cleveland) Inc |