DateProductReasonClassFirm
Aug 18, 2026 POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Doubl... Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device. Class II Olympus Corporation of the Americas
Aug 6, 2026 ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Number... The linear actuator and rotational lock mechanism components may experience a mechanical failure ... Class II IMRIS Imaging Inc
Aug 6, 2026 ORT200, operating room table, including the ORT200 Covered Table Assembly, Re... The linear actuator and rotational lock mechanism components may experience a mechanical failure ... Class II IMRIS Imaging Inc
Aug 4, 2026 Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digit... Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X... Class II Canon Medical System, USA, INC.
Jul 30, 2026 BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit ... After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or im... Class I Bard Access Systems, Inc.
Jul 29, 2026 healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative... Test manufactured with an expired ingredient which may present as either a false negative or posi... Class II Healthmark Industries Co., Inc.
Jul 29, 2026 healthmark ProChek-II, Model Number PT-313, PT-201-12 Test manufactured with an expired ingredient which may present as either a false negative or posi... Class II Healthmark Industries Co., Inc.
Jul 24, 2026 Ami HTX Momentary absence of On-Screen "X-Ray On" banner during run readiness check. Class II Spectral Instruments Inc
Jul 24, 2026 Lago X Momentary absence of On-Screen "X-Ray On" banner during run readiness check. Class II Spectral Instruments Inc
Jul 22, 2026 FUJIFILM Synapse PACS software v7.4.310; radiological image processing system Software issue may cause the ruler measurement to be inaccurate. Class II FUJIFILM Healthcare Americas Corporation
Jul 22, 2026 Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation Software issue may cause measurements to be displayed inaccurately. Class II Merge Healthcare, Inc.
Jul 20, 2026 PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addit... Class II GE Medical Systems Information Technologies Inc
Jul 20, 2026 Dentsply Sirona Capsule Mixer Model No. 5546068 Potential for electrical discharge from unit when unplugged. Class II SDI Limited
Jul 20, 2026 Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Soft... A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an i... Class II Midmark Corporation
Jul 20, 2026 FRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 570... Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addit... Class II GE Medical Systems Information Technologies Inc
Jul 20, 2026 Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software ... A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an i... Class II Midmark Corporation
Jul 20, 2026 PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addit... Class II GE Medical Systems Information Technologies Inc
Jul 20, 2026 PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addit... Class II GE Medical Systems Information Technologies Inc
Jul 20, 2026 Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF#... Due to a firmware defect, the portable organ transfer device datalogger may lose its configuratio... Class II Paragonix Technologies Inc.
Jul 17, 2026 Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number ... Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria... Class II Remel Europe Ltd.
Jul 16, 2026 Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 7220... Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degr... Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Jul 16, 2026 Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxoli... Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently con... Class II Straumann USA LLC
Jul 16, 2026 Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 7222... Two software related issues: 1) Potential image quality degradation that may lead to delay of t... Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Jul 16, 2026 FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)008514580051... Firm has identified mismatches between the outbreak strain and the reference strain used in the d... Class II Biofire Defense
Jul 15, 2026 Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5... a software defect in which the Parameter Package (PP) import logic overwrites all predefined cal... Class II Beckman Coulter Inc.
Jul 15, 2026 Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1... a software defect in which the Parameter Package (PP) import logic overwrites all predefined cal... Class II Beckman Coulter Inc.
Jul 15, 2026 BMC/REACTHEALTH Luna G3 APAP REF LG 3600 Retroactive reporting for 2024 rework conducted at the 3B/React Health warehouse Class I BMC Medical Co., Ltd.
Jul 14, 2026 CONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable Driver REF# C... Suture anchor may contain incorrect patient implant card which may result in incorrect device ide... Class II ConMed Corporation
Jul 14, 2026 Suture Cinch Long 1pk. Endoscopic tissue approximation device. Device may fail to deploy the cinch implant and/or fail to cut the suture as intended. Class II Boston Scientific Corporation
Jul 14, 2026 Onli 17cm Hydrophilic Urethral Intermittent Catheter, labeled as: 1) 82081... Hollister has identified that select lots of the catheters identified in the above table may exhi... Class II Hollister Ulc
Jul 14, 2026 OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device. Device may fail to deploy the cinch implant and/or fail to cut the suture as intended. Class II Boston Scientific Corporation
Jul 14, 2026 Cardinal Health Presource Kits: 1) BASIC PACK , Catalog N... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) SEPTO RHINO PACK , Catalog Nu... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) NEWNAN SMR PACK, Catalog Number: SEN21SPPMB Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2 , Catalog Nu... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) FAT GRAFT ADD KIT , Catalog Nu... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) KIT, MER NEURO BACK , Catalog N... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) HARMONIZED SPINE VCH PHC PACK , Catalog N... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 14, 2026 Cardinal Health Presource Kits: 1) ENT TRAY , Catalog Nu... Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected v... Class II Cardinal Health 200, LLC
Jul 13, 2026 Innercool Coolblue Console, Model STx; Thermal regulating system. Overheating of the 115V CPU board may cause the device to shut down. Class II Gentherm Medical, LLC
Jul 13, 2026 Blanketrol III, Model 233; Thermal regulating system. Overheating of the 115V CPU board may cause the device to shut down. Class II Gentherm Medical, LLC
Jul 13, 2026 MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; Loaner insulin pumps may have been cleaned with chemically incompatible products that could damag... Class II Medtronic MiniMed, Inc.
Jul 13, 2026 Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO... Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wi... Class II Medline Industries, LP
Jul 13, 2026 MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782; Loaner insulin pumps may have been cleaned with chemically incompatible products that could damag... Class II Medtronic MiniMed, Inc.
Jul 13, 2026 Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D... Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient resul... Class II Siemens Healthcare Diagnostics Products Ltd.
Jul 13, 2026 Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY... Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient resul... Class II Siemens Healthcare Diagnostics Products Ltd.
Jul 13, 2026 MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW Loaner insulin pumps may have been cleaned with chemically incompatible products that could damag... Class II Medtronic MiniMed, Inc.
Jul 13, 2026 Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product ... Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wi... Class II Medline Industries, LP
Jul 13, 2026 Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYS... Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the ... Class II Kico Knee Innovation Company
Jul 13, 2026 Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine As... Class II Siemens Healthcare Diagnostics, Inc.