| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Sep 21, 2022 | WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneury... | Detachment controller, of an aneurysm embolization system, has an out of specification tunnel ope... | Class II | MICROVENTION INC. |
| Sep 20, 2022 | Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 67... | Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems... | Class II | Medtronic MiniMed |
| Sep 20, 2022 | ARTIS pheno, Interventional fluoroscopic x-ray system, Model No. 10849000 | Siemens has become aware of three potential software issues with ARTIS pheno systems with softwar... | Class II | Siemens Medical Solutions USA, Inc |
| Sep 20, 2022 | STRATAFIX SPIRAL PGA-PCL UNI UD 20CM USP2-0(M3) S/A SH. indicated for use in ... | Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may re... | Class II | Ethicon, Inc. |
| Sep 20, 2022 | Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-711... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | Allen Life Assist Beach Chair, BCSSP-BEACH CHAIR , REF A-91501, #104002. For... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | C-Flex Polar Head Positnr-Device Only, SSP-NEUROSURGICAL HEADREST, REF A-7070... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104.... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devic... | Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instr... | Class II | Intuitive Surgical, Inc. |
| Sep 20, 2022 | Teleflex Medical RUSCH Pocket Pac i.c. Closed System Intermittent Catheter Ki... | Product may not be sterile | Class II | TELEFLEX LLC |
| Sep 20, 2022 | Simple Clamp, US/EURO Rails, UPPER EXTREMITY DISTRACTION CMPT, REF F-40018, #... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | Allen Standard Starburst, Mayfield/Doro, STANDARD SPINE POSITIONERS CMPT, R... | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly,... | Class II | Baxter Healthcare Corporation |
| Sep 20, 2022 | Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712... | Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems... | Class II | Medtronic MiniMed |
| Sep 20, 2022 | Vortex MP Port Titanium Low Profile Implanted Port with 5F attached Silicone ... | Printed Instructions for Use (IFU) not shipped with device. The Instructions for Use contain the ... | Class II | Angiodynamics, Inc. |
| Sep 19, 2022 | Atellica CH Lithium (Li), Siemens Material Number 11097535. in vitro diagnost... | Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the wors... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 19, 2022 | 125 series iGO2 Portable Oxygen Concentrator-indicated for the administratio... | An unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patie... | Class II | DeVilbiss Healthcare LLC |
| Sep 19, 2022 | Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro di... | Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the wors... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 16, 2022 | Prep-Fill Guarded Luer Connector (GLC): a. Part Number 57-400, Female Lo... | Blister package had an unsealed edge compromising sterility. | Class II | International Medical Industries, Inc. |
| Sep 16, 2022 | LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004 | Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop)... | Class II | Fresenius Kabi USA, LLC |
| Sep 16, 2022 | Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical a... | Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of acces... | Class II | Olympus Corporation of the Americas |
| Sep 15, 2022 | B. Braun CARESITE Port Access Kit w/Tegaderm, REF 375214, Central Venous (car... | The kits were incorrectly labeled with an extended expiration date of 02/21/2024 instead of 02/10... | Class II | Medical Action Industries, Inc. 306 |
| Sep 15, 2022 | CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per c... | The extension set has a high probability of potential failure which could lead to possible leakag... | Class II | B. Braun Medical, Inc. |
| Sep 15, 2022 | ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 uni... | The extension set has a high probability of failure which could lead to possible leakage or deta... | Class II | B. Braun Medical, Inc. |
| Sep 15, 2022 | BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog... | The affected product contains isobutylene which has recently demonstrated potential interference ... | Class II | Becton Dickinson & Company |
| Sep 15, 2022 | Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per ca... | The extension set has a high probability of failure which could lead to possible leakage or detac... | Class II | B. Braun Medical, Inc. |
| Sep 14, 2022 | Prismaflex M Set: M150 (product code 109990). For use in providing continuo... | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Ex... | Class II | Baxter Healthcare Corporation |
| Sep 14, 2022 | Disc Preparation instruments contained in the following products: CI-11050-0... | Due to manufacturing instruments which included additional features and adequate consideration wa... | Class II | Alphatec Spine, Inc. |
| Sep 14, 2022 | Prismaflex HF Set: HF1000 (product code 107140), HF1400 (product code 107142... | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Ex... | Class II | Baxter Healthcare Corporation |
| Sep 14, 2022 | Y-CONNECTOR: 6" Small Kink Resistant Tubing Bifurcated Connector /2 Male Lue... | Sterile extension sets were distributed without an expiration date | Class II | Avanos Medical, Inc. |
| Sep 14, 2022 | ARIES SARS-CoV-2 Assay, REF: 50-10047 | A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction ... | Class II | Luminex Corporation |
| Sep 14, 2022 | Prismaflex ST Set: ST60 (product code 107643), ST100 (product code 107636), ... | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Ex... | Class II | Baxter Healthcare Corporation |
| Sep 14, 2022 | Prismaflex M Set: M60 (product code 106696), M100 (product code 106697). Fo... | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Ex... | Class II | Baxter Healthcare Corporation |
| Sep 14, 2022 | Prismaflex HF20 Set (product code 109841). For use in providing continuous f... | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Ex... | Class II | Baxter Healthcare Corporation |
| Sep 14, 2022 | ARROW Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Midline Kit, 5.5 ... | Mislabeling: the corrugate label, lidstock, banner card, and peel sticker accompanying the affe... | Class II | ARROW INTERNATIONAL Inc. |
| Sep 13, 2022 | NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176 | Failure of ball stud components causing the gas strut on one side to detach from the canopy, maki... | Class II | Thor Photomedicine Ltd |
| Sep 13, 2022 | G041 POD, THOR NOVO GEN 2.0 XL whole body red light therapy bed, model S1177 | Failure of ball stud components causing the gas strut on one side to detach from the canopy, maki... | Class II | Thor Photomedicine Ltd |
| Sep 13, 2022 | NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184. | Failure of ball stud components causing the gas strut on one side to detach from the canopy, maki... | Class II | Thor Photomedicine Ltd |
| Sep 13, 2022 | NovoTHOR Gen 2.5 Regular whole body red light therapy bed, model S2183 | Failure of ball stud components causing the gas strut on one side to detach from the canopy, maki... | Class II | Thor Photomedicine Ltd |
| Sep 12, 2022 | Orthex Large Bone Shoulder Bolt, model no. AS-17 | Bolt breakage on the head or threaded portion, which may cause loss of strut continuity with fram... | Class II | OrthoPediatrics Corp |
| Sep 9, 2022 | MEDLINE PHERIPHERAL BLOOD CULTURE DRAW, Reorder #: DYNDH1561 | Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration d... | Class II | MEDLINE INDUSTRIES, LP - Northfield |
| Sep 9, 2022 | MEDLINE PRE-OP CHG BATHING KIT Reorder #: P164756 | Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration d... | Class II | MEDLINE INDUSTRIES, LP - Northfield |
| Sep 9, 2022 | MEDLINE BLOOD CLUTURE KIT, Reorder #: DYNDH1592 | Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration d... | Class II | MEDLINE INDUSTRIES, LP - Northfield |
| Sep 9, 2022 | MEDLINE 24 BLOOD CLUTURE KIT, Reorder #: DYNDH1632 | Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration d... | Class II | MEDLINE INDUSTRIES, LP - Northfield |
| Sep 9, 2022 | MEDLINE ADLT BLOOD CLUTURE COLLECT KIT, Reorder #: DYNDH1245 | Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration d... | Class II | MEDLINE INDUSTRIES, LP - Northfield |
| Sep 8, 2022 | Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip N... | An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked... | Class II | Abbott Vascular |
| Sep 8, 2022 | Full Kit Pulset 3cc Syringe 23g x 1 Safety Tip 25u BH, REF Number 3373-91 | The kits are being recalled due to a lack of sterility of a spare needle contained within the kit... | Class II | SunMed Holdings, LLC |
| Sep 8, 2022 | Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G... | An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked... | Class II | Abbott Vascular |
| Sep 8, 2022 | Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G... | An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked... | Class II | Abbott Vascular |
| Sep 8, 2022 | cobas 5800 instrument, Material No. 08707464001. Used for automated Polymeras... | Reported false positive and invalid results on the affected devices due to anomalous baselines. T... | Class II | Roche Molecular Systems, Inc. |
| Sep 7, 2022 | THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S... | Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irr... | Class II | Johnson & Johnson Surgical Vision, Inc. |