Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Ad...
FDA Device Recall #Z-3188-2026 - Class II, July 22, 2026
Recall Summary
| Recall Number | Z-3188-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | July 22, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Boston Scientific Corporation |
| Location | Marlborough, MA |
| Product Type | Devices |
| Quantity | 982 units |
Product Description
Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0
Reason for Recall
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.
Lot / Code Information
All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)
Other Recalls from Boston Scientific Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-3214-2026 | Class I | Imager II Angiographic Catheters M001314001 ... | Aug 24, 2026 |
| Z-3082-2026 | Class II | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardia... | Aug 6, 2026 |
| Z-3083-2026 | Class II | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardia... | Aug 6, 2026 |
| Z-3081-2026 | Class II | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardia... | Aug 6, 2026 |
| Z-2962-2026 | Class II | Suture Cinch Long 1pk. Endoscopic tissue app... | Jul 14, 2026 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery, the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.