Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD1...
FDA Device Recall #Z-0717-2018 - Class II, December 20, 2017
Recall Summary
| Recall Number | Z-0717-2018 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 20, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips Electronics North America Corporation |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 1486 |
Product Description
Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper FD20, Allura Xper FD20/10, Allura Xper FD20/15 and Allura Xper FD10/10 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and ElectroPhysiology. Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room (OR Table)
Reason for Recall
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
Distribution Pattern
Worldwide Distribution - US Nationwide
Lot / Code Information
All Allura Xper FD and and Xper FD ORT manufactured from July 2003 - May 2017 with system Codes: 722001 722002 722003 722010 722026 722005 722011 722027 722006 722012 722028 722008 722013 722015 722023 722035 722020 722029 722038 722039 722058 722059
Other Recalls from Philips Electronics North America Co...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0614-2019 | Class II | Xper Flex Cardio Physiomonitoring System, Softw... | Nov 15, 2018 |
| Z-0592-2019 | Class II | IntelliVue MX40 Patient Monitor, Model Nos. 865... | Oct 10, 2018 |
| Z-3227-2018 | Class II | Allura Xper F010, System Code 722003 Vascul... | Jul 16, 2018 |
| Z-3218-2018 | Class II | INTEGRIS SUITE, System Code 722199 Vascular... | Jul 16, 2018 |
| Z-3215-2018 | Class II | INTEGRIS CV, System Code 722030 Vascular, c... | Jul 16, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.